Radio Shack 22-811 Owner'S Manual Download Page 6

6

Note:  With no resistance connected 
across the test leads (meaning 
resistance is infinite), 

E$<

E$<

E$<

E$<

 appears when 

you set the meter to measure 
resistance. This is normal.

4. Touch the test leads across the circuit 

you want to measure, or remove one of 
the leads of the component you want to 
measure from its circuit and touch the 
test leads across the component. If the 
meter is set to automatic range control, it 
automatically moves to the proper 
range.

If you set the meter to use manual range, 
repeatedly press 

RANGE 

to set manual-range 

control and change the range (if necessary).

Notes: 

• If you are measuring resistance of about 

1M

 or more, the display might take a 

few seconds to stabilize. This is normal.

• As with the voltage range, use the 

measuring units that appear on the 
display to determine the current 
resistance range. If only 

 appears, the 

values of the measurements are in 
ohms. If 

AAAA

 and 

 appear, the meter is 

measuring kilohms (1 kilohm = 1000

). 

If 

C

C

C

C

 and 

 appear, the meter is 

measuring megohms (1 megohm = 
1,000,000 

).

Hint: If you want to accurately measure a 
very small resistance, you can view the 
resistance of the meter’s test leads, then 
subtract that resistance from the measured 
value. To measure the meter’s internal 
resistance, simply touch the end of the test 
leads together. The meter selects the 400

 

scale and displays its internal resistance. 
You can also use the meter’s relative 
function to do this (see “Using Relative 
Measurement” on Page 4).

/

'#574+0)

"

%

#2#%+6#0%'

The capacitance measuring circuit in your 
meter charges a connected capacitor to a 
specific voltage level, then discharges the 
capacitor to a lower voltage. The meter 
measures the amount of time takes to 
discharge the capacitor.

Caution: Do not connect the test leads to a 
source of voltage with the function selector 
set to 

. This could damage the meter or 

the circuit being tested.

1. Remove all power from the circuit under 

test and discharge all capacitors.

2. Rotate the function selector to 

.

3. Set the meter to measure capacitance 

by repeatedly pressing 

SELECT

 until 

nF

 

or 

µ

F

 appears.

4. Attach the red test lead to the positive 

side of the capacitor and attach the 
black test lead to the negative side of the 
capacitor. Or remove one of the leads of 
the capacitor you want to measure from 
its circuit and connect the test leads to 
the capacitor.

Note:  Electrolytic capacitors have 
positive and negative terminals. Be sure 
to connect each test lead to the correct 
terminal to match the polarity of these 
capacitors.

5. Read the measured value on the display.

Notes:  

• The voltage applied across electrolytic 

capacitors will affect their measured 
values. That is, a meas-urement taken 
with a low voltage will be different and 
lower than that taken with a voltage that 
approaches the capacitor’s voltage 
rating. Since this meter cannot use high 
voltage to set the electrolyte. It cannot 
measure the absolute value of 
capacitance.

• Low-value capacitors might match or be 

close to the actual input capacitance of 
your meter. To measure low capacitance 
values, use the meter’s relative feature 
to adjust the meter for the meter’s own 
capacitance (see “Using Relative 
Measurement” on Page 4).

• The accuracy of capacitance 

measurement depends on the 
measurement method and also differs 
with different types of capacitors. The 
meter’s measurement is for reference 
only.

%

*'%-+0)

"

%

106+07+6;

Caution: Do not connect the test leads to a 
source of voltage with the selector set to 

/

. This could damage the meter or the 

circuit being tested.

You can use the meter to check for shorted 
or open electrical circuits.

1. Remove all power from the circuit under 

test and discharge all capacitors.

2. Rotate the function selector to 

/

.

3. To select the continuity function, 

repeatedly press 

SELECT

 until 

 

appears at the top of the display.

4. Touch the test leads across the circuit 

you want to measure. 

Shrt

 

appears and 

the buzzer sounds if the circuit 
resistance is less than about 50 ohms 
(meaning the circuit is continuous or 
shorted).  

Open

 appears and the meter’s 

buzzer does not sound if the circuit 
resistance is greater than about 50 

ohms (meaning the circuit is not 
continuous).

%

*'%-+0)

"

&

+1&'5

This procedure lets you check diodes, 
transistors, and other semiconductors for 
opens, shorts, and normal operation. It also 
lets you determine the forward voltage and 
polarity for diodes. (This is handy when you 
need to match a diode). You can also check 
LEDs using this procedure.

Caution:  Do not connect the test leads to a 
source of voltage with the function selector 
set to 

/

.  This could damage the meter 

or the circuit being tested.

1. Remove all power from the circuit under 

test and discharge all capacitors.

2. Rotate the function selector to 

/

.

3. To select the diode function, repeatedly 

press 

SELECT

 until 

 appears on the 

top of the display.

4. Connect the test leads to the device you 

want to check, or remove one of the 
leads of the component you want to 
measure from its circuit and connect the 
test leads to the component. Note the 
first reading.

5. Reverse the test leads and note the 

second reading.

If one reading shows a value and the other is 
overrange (

E$<

E$<

E$<

E$<

 appears) the device is good. If 

E$<

E$<

E$<

E$<

 appears during both readings, the device 

is open. If both values are very small or zero, 
the device is shorted.

Notes: 

• When you test a silicon-type 

semiconductor, the values might vary 
depending on the temperature.

• The values that appear during a diode 

check show the actual forward voltage 
(2.0V max). If the voltage exceeds 2.0V, 

E$<

E$<

E$<

E$<

 appears. This means the diode 

check cannot be made using this meter.

%JGEMKPI

"

&KQFG

"

2QNCTKV[

Many diodes have a stripe or mark on one 
side. The marked side of the diode indicates 
the diode’s cathode or negative (

) side. The 

other side is the anode or positive (+) side.

If a diode is not marked, you can use your 
meter to check the diode’s polarity. As you 
follow the steps under “Checking Diodes” on 
Page 6, connect the red test lead to one side, 
connect the black test lead to the other side, 
then measure and note the voltage. Then 
reverse the test leads and measure and note 
the second reading. The side of the diode 
where the meter shows a higher voltage 
using the red test lead is the anode (+) side.

Summary of Contents for 22-811

Page 1: ...d the stringent safety tests required by Underwriters Laboratories Note The UL mark does not indicate that this product has been evaluated by Underwriters Laboratories for the accuracy of its readings Latest IC and Display Technology ensures reliability accuracy stability and ease of operation Overload and Transient Protection helps protect the meter from accidental overload in most ranges Note Yo...

Page 2: ...st Digit Note Accuracy with film capacitor or better specified from 9 5 of full scale to full scale except 4 0nF range from 0 5nF to full scale FREQUENCY Frequency Range 10Hz to 4MHz 400Hz 4k 40k 400kHz 4MHz 0 1 of Reading 4 in Last Digit Sensitivity 400Hz 4k 40kHz 50 mVrms 400kHz 100 mVrms 4MHz 350 mVrms DUTY CYCLE 10Hz to 100kHz 10 90 Square wave at 5V 0V 2 digits kHz 2 in last digit Note Duty C...

Page 3: ...F and 83 F 18 C and 28 C with a maximum of 75 relative humidity 6740 0 6 6 4 1011 1 6 56 0 6 52 To turn on the meter rotate the function selector to any function except OFF To turn off the meter rotate the function selector to OFF Note If the function selector is not set to OFF and nothing appears on the display the meter might be in its auto power shut off mode Press any button to turn on the met...

Page 4: ...e test leads connected to a component hold down REL Rel appears The meter is set to manual range mode 4 Use the meter to make a measurement Notes If the reference value you measured in Step 2 is greater than the value of the measurement you made in Step 4 the meter shows a negative value This is not a malfunction E E E E appears if you measure values outside of the meter s currently set range The ...

Page 5: ...circuit Doing so could present a dangerous shock hazard to you and could also damage the meter 574 0 1 744 06 To measure AC or DC current you must break the circuit and connect the test leads to two circuit connection points The connection must be in series with the circuit under test WARNING Do not apply voltage directly across terminals You must connect the meter in series with the circuit Cauti...

Page 6: ... capacitance of your meter To measure low capacitance values use the meter s relative feature to adjust the meter for the meter s own capacitance see Using Relative Measurement on Page 4 The accuracy of capacitance measurement depends on the measurement method and also differs with different types of capacitors The meter s measurement is for reference only 0 106 07 6 Caution Do not connect the tes...

Page 7: ...er and the top cabinet together Do not try to remove the back cover 5 To remove the fuse gently pull the red ribbon holding it The fuse pops out 6 If the fuse is blown discard it and save the ribbon Then remove the spare fuse from the plastic holder and insert it into the fuse holder through the loop of the attached ribbon 7 Replace the back cover and secure it with the screws 8 Install the batter...

Reviews: