Radiation Alert Inspector EXP+ User Manual Download Page 9

9

5  Maintenance

The Inspector

+

requires regular calibration and careful handling to assure good measurements.  Use the following

guidelines to maintain the Inspector

+

properly.

Calibration

The Inspector

+

should be calibrated as often as your regulations require, or in any case, at least once a year.  The best

way to calibrate is using a calibrated source.  If no source is available, it is possible to calibrate electronically using a pulse
generator.
The standard radionuclide for calibration is Cesium-137.  A certified calibration source should be used.  To calibrate the
Inspector

+

for another radionuclide, you must use a calibrated source for that radionuclide or the appropriate conversion

factor referenced to Cs-137.

CAUTION:

Errors can occur when using low level sources or background to set 

CAL factor.  In the Calibration

mode, the smallest increment which can be adjusted is .010, which prevents fine adjustment of the CAL factor.

Calibrating Using a Source

1.  Place the Inspector

+

or Inspector EXP

+

probe at a distance from the source that corresponds to a 50 mR/hr field with

the detector window facing the source. 

2.  Set the Inspector

+

mode switch to mR/hr.

3.  Turn the Inspector

+

on.

4.  Open the source and record 20 consecutive readings.
5. Close the source.
6. Calculate the average of the readings and record.

a) If the average is ± 10% of 50 mR/hr, go to Step 7.

b) If the average is not ± 10% of 50 mR/hr, go to Step 10.

7.  Place the Inspector

+

or the Inspector EXP

+

probe at a distance from the source that corresponds to a 5 mR/hr field

with the detector window facing the source.

8.  Repeat Steps 2 - 5.
9.  Calculate the average of the readings and record.

a) If the average is ± 10% of 5 mR/hr, the calibration procedure is complete.

b) If the average is 

not

± 10% of 5 mR/hr, go to Step 10.

10. Turn off the AUDIO in order to hear the count down timer sound.
11. Turn off the Inspector

+

.  Hold down the “-” button on the end panel and turn on the Inspector

+

.  The display shows

CAL, and the Inspector

+

counts down for 15 seconds, chirping each second. This delay gives you a chance to move

out of the field and then expose the source. At the end of the 15 seconds, the Inspector

+

beeps.

12. The Inspector

+

now collects data for 30 seconds, chirping every 2 seconds, with CAL and the hourglass icon flashing.

At the end of the 30 seconds, it beeps. The display shows CAL and SET is flashing.

13. Close the source.
14. Press the “+” and “-” buttons on the Inspector

+

to adjust the reading to what it should be.

15. When the reading is correct, press the Set button. The new calibration factor is displayed for several seconds, then

the Inspector

+

beeps and resumes normal operation.

16. Record the new calibration factor.
17. Place the Inspector

+

or the Inspector EXP

+

probe at a distance from the source that corresponds to a 5 mR/hr field

with the detector window facing the source.

18. Repeat Steps 2 - 5.
19. Calculate the average of the readings and record.|

a) If the average is ± 10% of 5 mR/hr, the calibration procedure is complete.

b) If the average is not ± 10% of 5 mR/hr, repeat steps 11 - 16 and go to step 20.

20. Calculate the average of the calibration factor for 50 mR/hr and the calibration factor for 5 mR/hr.
21. Turn the Inspector

+

off. Hold down the “+” button while turning the Inspector

+

on.  The numeric display will show a

single number.

23. Press the “+” or “-” button until 7 is shown on the numeric display.
24. Push the Set button.
25. The calibration factor is displayed.  Press the “+” or “-” buttons to adjust the calibration factor to the average

calibration factor calculated in Step 20.

26. Push the Set button to enter the new setting and resume normal operation.

The calibration factor is set to 100 (percent) at the factory. If you change the reading, for example, to 20% higher than the
factory reading, the new calibration factor would be 120. The current calibration factor is displayed during the system

Summary of Contents for Inspector EXP+

Page 1: ...User Manual...

Page 2: ...lity Menu 7 Options 8 4 Common Procedures 8 Establishing the Background Count 8 Environmental Area Monitoring 8 Checking for Surface Contamination 8 5 Maintenance 9 Calibration 9 Calibrating Using a S...

Page 3: ...spector in temperatures over 100 F 38 C or in direct sunlight for extended periods of time Do not get the Inspector wet Water can damage the circuitry and the mica surface of the Geiger tube Do not pu...

Page 4: ...nel Each switch has three settings which are described below Mode Switch Figure 1 2 mR hr Sv hr The numeric display shows the current radiation level in milliroentgens per hour or when SI units are us...

Page 5: ...et through the plastic case to penetrate the mica surface of the tube The small radiation symbol Figure 1 7 on the front label indicates the center of the Geiger tube External For Inspector EXP only T...

Page 6: ...ility Menu detailed in Chapter 3 Refer to the following table After 30 second start up the reading will be based if instrument is detecting on an average of the previous 100 CPS 30 seconds 6000 CPM or...

Page 7: ...e Using Dose Rate Modes While Timer is On Dose rate modes can be used while the timer is on In any dose rate mode the hour glass indicator will continue to flash during a timed period At the end of th...

Page 8: ...e information on using the timer see Taking a Timed Count in Chapter 3 A ten minute average is moderately accurate You can repeat it several times and see how close the averages are To establish a mor...

Page 9: ...nt down timer sound 11 Turn off the Inspector Hold down the button on the end panel and turn on the Inspector The display shows CAL and the Inspector counts down for 15 seconds chirping each second Th...

Page 10: ...L and the hourglass icon flashing At the end of the 30 seconds it beeps The display shows CAL and SET is flashing 7 Press the and buttons to adjust the reading to what it should be When the reading is...

Page 11: ...ta radiation A beta particle consists of an electron emitted from an atom It has more mass and less energy than a gamma ray so it doesn t penetrate matter as deeply as gamma and X rays Alpha radiation...

Page 12: ...r 3 Operation Autoranging Display update CAL Factor Range 001 to 199 Timer Can set 1 10 minute sampling periods in one minute increments 10 50 minute sampling periods in 10 minute increments and 1 24...

Page 13: ...or carrying the instrument Appendix C Inspector EXP Carrying Case The reinforced protective probe cover prevents damage to the fragile window of the detector The front flap lifts out of the way to gai...

Page 14: ...m Observer Software The Observer software runs on a Windows platform and can be used with the Inspector Digilert 50 and Geiger Radiation Monitors As an option any of our Radiation Alert instruments ca...

Page 15: ...onform to the warranty at any time while this warranty is effective the Warrantor will repair the defect and return the instrument to you prepaid without charge for parts or labor NOTE While the produ...

Page 16: ...e company serial no Inside battery compartment or rear label address City state zip code 4 calibrations per year circle 1 2 3 4 phone number date placed in service Mail to Attn Robert Russell S E Inte...

Reviews: