  3.2 Forehead mode

: Press the “Forehead” button, you can get the body temperature by forehead   

                                  measurement after hearing two beep sounds. (see figure 2) 

 

  3.3

 

Surface mode

:

 Press and hold the “ON/MEM” button, and press the “Forehead” button one time

,     

                              you can see the 

  icon on your LCD display(see figure 3).   

  In this mode, press the “Forehead” button to get the target surface temperature. 

  3.4 Stopwatch

:

 Press and hold the “ON/MEM” button, and press the “Forehead” button twice

, you can   

see the 

  icon on your LCD display(see figure 4). 

In this mode, press the “Forehead” button to count by 0.01 second to 3 minutes.   

4. Temperature taking:     

4.1 Ear temperature measuring, points for attention:   

Remarks: a. It is recommended that you measure 3 times with the same ear. If the 3 measurements are different, select the highest temperature. 

b. To avoid the risk of cross contamination, please clean the probe according to “

Care and cleaning

” section after each use.

 

                                      c. The different temperature varies in healthy persons between different parts of the body can be between 0.2~1°C 

                                    d. Clinical repeatability: 0.23°C (<1 year old),0.22°C (1~5 years old),0.21°C (>5 years old) 

 

4.1.1 Gently pull the ear back to straighten the ear canal (Fig.4.1.1) and snugly position the probe into the ear canal, aiming towards 

the membrane of the eardrum to obtain an accurate reading (Fig.4.1.2).   

4.1.2  Press  and  hold  the  “Ear”  button  until  you  hear  a  beep  sound.  Remove  the  probe  from  ear  and  read  the  temperature 

measurement on the LCD.

 

 

4.2 Forehead temperature measuring, points for attention:

   

Remarks: a. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its “oral-equivalent” Value. 

b. Before the measurement, please stay in a stable environment for 5 mins and avoid exercise, bath for 30 mins. 
c. Remember to keep the temple area clean and away from sweat, cosmetics and scar while taking temperature. 

 

4.2.1 The temporal artery is connected to the heart via the carotid artery. It is designed to measure the skin surface around 

temporal artery, a major artery of the head (Fig.4.2.1).

 

4.2.2 Attach the thermometer to the required measurement location on the forehead (Fig. 4.2.2). (You can choose left or 

right temple.)   

4.2.3 Press the “Forehead” button, and gently scan around the temple area (Fig. 4.2.3). While scanning, you will hear a 

beep sound, which indicates you the newest measurement is taking place. Measurement has been completed after 
two short beeps are heard and the forehead icon stops flashing. The time consuming for measurement might be 
between 5~8 sec (up to 30sec). It depends on how much time the device needs to get the correct forehead 
temperature.

 

4.3 Measuring temperature under surface mode:   

4.3.1 When you press the “Forehead” button, you will get the real time temperature immediately. If you press and hold the “Forehead” button, the reading of 

measurement will be continuous updated.   

4.3.2 Applications include temperature measurements for Water, Milk, Cloth, Skin or other object. 

* Note: This mode shows the actual, unadjusted surface temperatures, which is different from the body temperature. 

 

4.4 Get the room temperature 

      4.4.1 Setting “Real time clock” first to enable this function. After power off, room temperature will be shown on the screen with 

  icon. 

4.4.2 For room temperature, the thermometer should be placed on the table and avoid the direct sunshine or air conditioner flow on.   
4.4.3 The room temperature can be reference after 15min later. 

5. After measurement: 

5.1 Power off: Device will automatically shut off if left idle for more than 1 minute to extend battery life. The LCD will only show the date, time and room temperature. 
5.2 Clean the probe after each use to ensure an accurate reading and avoid cross contamination.    (See the section of Care and Cleaning for details.) 
5.3 Remember to put on the Probe Cap when not in use.   

 

Use of the Stopwatch

 

1.

 

Press and hold the “ON/MEM” button, and press the “Forehead” button twice for “stopwatch” mode. 

2.

 

For start or stop counting, press the “Forehead” button; for reset counting, press the “ON/MEM” button.   

3.

 

Press and hold the “ON/MEM” button, and press the “Forehead” button to enter the forehead mode. 

4.

 

How to measure the heartbeat? If you count out the number of beating is 18 in 15sec, then you can calculate the heartbeat of patient is 72 times in 1min. 

 

Important Notes 

Care and cleaning

 

T

he probe is the most delicate part of the thermometer. Use with care when cleaning the lens to avoid damage.   

a. After the measurement, please use the cotton swab with the Alcohol (70% concentration) to clean the lens( on the inside of the probe). 
b. Allow the probe to fully dry for at least 1 minute. 
c.

 

Storage temperature range: It should be stored at room temperature between -20~+50°C , RH

85%   

d. Keep the unit dry and away from any liquids and direct sunlight.   
e. The Probe should not be submerged into liquids. 

* Note: Please check the device if damaged once it falls. If you can’t make sure of it ,please send the complete device to       

the nearest retailer for recalibration.   

 

  Holding the thermometer too long may cause a higher ambient temperature reading of the probe.   

This could make the body temperature measurement lower than usual. 

Figure 4.2.3 

Figure 4.2.2 

Figure 4.2.1 

Figure 4.1.1 

Figure 4.1.2 