Figure 4.2.3
Figure 4.2.2
Figure 4.2.1
3.3 Surface mode
:
After power on, press and hold the “ON/MEM” button, and press the “SCAN” button one time for “Surface” mode
to see
icon on your
LCD display. In this mode, you can get the target surface temperature. (see figure 3)
4. Temperature taking:
4.1 Ear temperature measuring, points for attention:
Remarks:
a. It is recommended that you measure 3 times with the same ear. If the 3 measurements are different, select the highest temperature.
b. To avoid the risk of cross contamination, please clean the probe according to “
Care and cleaning
” section after each use.
c. The different temperature varies in healthy persons between different parts of the body can be between 0.2~1°C
d. Clinical repeatability: 0.23°C (<1 year old), 0.22°C (1~5 years old), 0.21°C (>5 years old)
** When you get a new temperature reading in ear mode, the last reading will be shown on the screen (at top right corner) with the
icon.**
4.2 Forehead temperature measuring, points for attention:
Remarks:
a. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its “oral-equivalent” Value.
b. Before the measurement, please stay in a stable environment for 5mins and avoid exercise, bath for 30mins.
c. Remember to keep the temple area clean and away from sweat, cosmetics and scar while taking temperature.
4.2.1 The temporal artery is connected to the heart via the carotid artery. It is designed to measure the skin surface around
temporal artery, a major artery of the head (Fig.4.2.1).
4.2.2 Attach the thermometer to the required measurement location on the forehead (Fig. 4.2.2). (You can choose left or
right temple.)
4.2.3 Press the “SCAN” button, and gently scan around the temple area (Fig. 4.2.3). While scanning, you will hear a beep
sound, which indicates you the newest measurement is taking place. Measurement has been completed after two
short beeps are heard and the forehead icon stops flashing. The time consuming for measurement might be between
5~8 sec (up to 30sec). It depends on how much time the device needs to get the correct forehead temperature.
4.3 Measuring temperature under surface mode:
4.3.1 When you press the “SCAN” button, you will get the real time temperature immediately. If you press and hold the “SCAN” button, the reading of measurement
will be continuous updated.
4.3.2 Applications include temperature measurements for Water, Milk, Cloth, Skin or other object.
* Note: This mode shows the actual, unadjusted surface temperatures, which is different from the body temperature.
4.4 Get the room temperature
4.4.1 For room temperature, the thermometer should be placed on the table and avoid the direct sunshine or air conditioner flow on.
4.4.2 When in power off, room temperature will always be shown on the screen with
icon.
4.4.3 The room temperature can be reference after 15min later.
5. After measurement:
5.1 Power off: Device will automatically shut off if left idle for more than 1 minute to extend battery life. The LCD will only show the room temperature.
5.2 Clean the probe after each use to ensure an accurate reading and avoid cross contamination. (See the section of Care and Cleaning for details.)
5.3 After measurement, please push back the raised brim on the sliding panel on the above of LCD screen for protecting the probe.
Important Notes
Care and cleaning
T
he probe is the most delicate part of the thermometer. Use with care when cleaning the lens to avoid damage.
a. After the measurement, please use the cotton swab with the Alcohol (70% concentration) to clean the lens( on the inside of the probe).
b. Allow the probe to fully dry for at least 1 minute.
c. Storage temperature range: It should be stored at room temperature between -20~+50°C, RH
≦
85%
d. Keep the unit dry and away from any liquids and direct sunlight.
e. The Probe should not be submerged into liquids.
* Note: Please check the device if damaged once it falls. If you can’t make sure of it ,please send the complete device to
the nearest retailer for recalibration.
Holding the thermometer too long may cause a higher ambient temperature reading of the probe.
This could make the body temperature measurement lower than usual.
Battery replacement
○
1
Open the battery cover: First, make sure the mode of thermometer is ”forehead mode”. Insert a coin to twist it clockwise, the battery is
located under the twist cover at the rear of the device.
○
2
Hold the device and flip the battery out with a small screwdriver
○
3
Insert the new battery under the metal hook on the left side and press the right side of the battery down until you hear a click
○
4
Replace the battery cover
Keep the battery away from children.(This device is supplied with one lithium cell (CR2032x1).)
The positive (+) side up and the negative (-) side pointed down.
Disposal of the used batteries: please comply with the local regulation.
4.1.1 Gently pull the ear back to straighten the ear canal (Fig.4.1.1)
and snugly position the probe into the ear canal, aiming towards
the membrane of the eardrum to obtain an accurate reading
(Fig.4.1.2).
4.1.2 Press and hold the “SCAN” button until you hear a beep sound.
Remove the probe from ear and read the temperature
measurement on the LCD.
(Rear Cover)
Figure. 4.1.1
Figure 4.1.2