background image

TH90K(E) Multi-function Infrared Thermometer 

Functions 

Room Temperature

 

Suitable ambient temperature is important for the baby and patient, the thermometer always help you recognize the 
room temperature. 

 When in power off, room temperature will always be shown on the screen with 

  icon. 

 Please see the “Use of the thermometer” section to learn how to get the correct room temperature. 

 How to turn off Room Temperature: When in power on, press and hold the “ON/MEM” button for 5 sec. until   

OFF”

 will be shown on the screen to release the button. 

 

Ear Temperature 

The  thermometer  has  been  designed  for  practical  use.  It’s  not  meant  to  replace  a  visit  to  the  doctor.  Please  also 
remember to compare the measurement result to your regular body temperature. 

 Please see the “Use of the thermometer” section to learn how to measure the body temperature. 

 

Forehead Temperature 

The  thermometer  has  been  designed  for  practical  use.  It’s  not  meant  to  replace  a  visit  to  the  doctor.  Please  also 
remember to compare the measurement result to your regular body temperature. 

 Please see the “Use of the thermometer” section to learn how to measure the body temperature. 

 

Surface Mode 

The surface mode shows the actual, unadjusted surface temperatures, which is different from the body temperature. 
It can help you to monitor if the object temperature is suitable for the baby or patient, for example the baby’s milk. 

 Please see the “Use of the thermometer” section to learn how to measure the object temperature. 

 

Fever Alarm

 

If thermometer detects a body temperature

 

37.5°C, there will be a long beep sound followed by three short beep 

sound to warn the user for potential fever. 

 

Last Reading 

When  you  get  a  new  temperature  reading  in  ear  mode,  the  last  reading  will  be  shown  on  the  screen  (at  top  right 
corner) with the 

  icon. 

 

Memory Locations

 

There are total 25 sets memories for ear measurements.   

 When in power on, Press the “ON/MEM” button to see the temperature stored with 

  icon.   

 

°C / °F Switch

 

  In  “power  off”  mode  (The  LCD  only  shows  the  room  temperature).  Press  and  hold  the  “SCAN”  button,  then 

immediately press and release the “ON/MEM” button. After 3 seconds, icon “°C” will be switched to icon “°F”. (You 
may use the same way to switch back to “°C”) 

 

 

 

 

 

Use of the thermometer

 

1. Always make sure the probe is clean, and without damage.   
2. Power on:    Press “ON/MEM” button.   

3. Mode selection:   
3.1

 

Ear mode

: Press the “ON/MEM” button, push up the raised brim on the   

  sliding panel on the above of LCD screen(see figure 4)

 until   

hearing a click sound

 ,the mode of thermometer is ear mode.   

In this mode, 

please press the “SCAN” button (see figure 5)

 

to get the body temperature by ear measurement.(see figure 1) 

 
3.2 Forehead mode

: Press the “ON/MEM” button , push back the raised brim                                                   

on the sliding panel on the above of LCD screen(see figure 6), 
the mode of thermometer is forehead mode.   
In this mode, you can get the body temperature by forehead   
measurement.(see figure 2) 

 

 

 

Specifications 

 

Temperature measurement range:   

Ear/Forehead mode: 34~42.2°C (93.2~108°F), Surface mode: -22~80°C (-7.6~176°F) 

 

Operating temperature range: 10~40°C (50~104°F) 

 

Storage temperature Range: It should be stored at room temperature between -20~+50°C, 
RH

85% 

Transportation temperature shall be less than 70°C, RH

95% 

 

Comply with ASTM E1965-98, EN12470-5:2003 Clinical thermometers-Part   
5:Performance of infra-red ear thermometers(with maximum device), 
IEC/EN60601-1-2(EMC), IEC/EN60601-1(Safety) standards. 

 

Accuracy: 

    For ear mode: +/-0.2°C (0.4°F) within 35.5~42°C (95.9~107.6°F), 

+/-0.3°C (0.5°F) for other range.   

For forehead mode: +/-0.2°C (0.4°F) within 36~39°C (96.8~102°F), 

+/-0.3°C (0.5°F) for other range.   

For surface mode: +/-0.3°C (0.5°F) within 22~42.2°C (71.6~108°F),   

    /-4% or +/-2°C (4°F) whichever is greater. 

 

Fever alarm & Memory under “ Ear/Forehead mode ” 

 

°C / °F Switch mode 

 

Battery: one lithium cell battery (CR2032 *1pcs) 

 

This thermometer converts the ear/forehead temperature to display its “oral equivalent.” 

 

:LED Backlight (optional) 

     

LED Back

l

ight will automatically turned on after measurement, and automatically turned 

      off after 5 seconds 

 

Do not use this device for Live & Dead decision or Safety related applications.

 

Please 

consult with doctor if you have health concerns. 

 

There is no gender and age limitation for using infrared thermometer.   

  This is not an AP

 

or APG product.   

 

 

 

 

 

 

Figure 1              Figure 2 

     

Figure 3 

Figure 6

 

Figure 4

 

Ear Mode

 

 

Forehead Mode

 

Figure 5

 

Reviews: