background image

 

 

This thermometer 

is an adjusted mode thermometer that

 converts the ear temperature to display its “oral equivalent.”   

(according to the result of the clinical evaluation to get the offset value) 

 

Enclosure Rating: IP22 

 

Dimensions: 141.4 x 40.4 x 43.5 mm 

 

Weight: 67.3 grams including battery 

 

Battery life: around 3,000 continuous readings. 

 

Expected Service Life: 4 years 

 

The device should not submerge into any liquids and expose it to direct moisture. 

 

There is no gender and age limitation for using infrared thermometer.   

  This is not an AP

 

or APG product. 

 

Troubleshooting: 

Error Message 

Problem 

Solution 

 

Device stabilization is in process.

 

Wait until   stops flashing.

 

 

Battery is low and no more measurements are 
possible.

 

Replace the battery.

 

 

Measurement before device stabilization.

 

Wait for “Er1” to disappear. 

 

The ambient temperature is not within the normal 
temperature range between 10

 o

C and 40

 o

C (50

 o

~104

 o

F). 

Allow the thermometer to rest in a room at least 15 
minutes at room temperature: 10

 o

C and 40

 o

C (50

 o

~104

 o

F). 

 

Error 5~9, the system is not functioning properly. 

Unload the battery, wait for 1 minute and repower it. 
If the message reappears, contact the retailer for 
service. 

 

Temperature taken is higher than 42.2

 o

C (108.0

 o

F). 

Check the integrity of the probe cover and take a new 
temperature measurement. 

 

Temperature taken is lower than 34

 o

C (93.2

 o

F). 

Make sure the probe cover is clean and take a new 
temperature measurement. 

 

Device can not be powered on to the ready stage. 

Change with a new battery. 

Warranty: 12 months   
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)

 

 

Ex.SN:E512A000001, the first “E” is External, the second number ”5” is the last number of manufacture year, the third and the 
fourth number “12” is the manufacture month, the others is the serial number.

 

Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of 

temperature measurements or unexpected events at any time, please contact the dealers or nearest service 
address. 

Please report to the manufacturer and the competent authority of the Member State in which you are established 
about any serious incident that has occurred in relation to this device.

 

 

Warning: No modification of this equipment is allowed. 

 

Please read the instructions for use       

  BF type applied part 

 

 

IMPORTER 
Company Name?    

Company Name? 
General Information: Address? Website? 

Radiant Innovation Inc.                                         Medical Technology Promedt Consulting Gmb 
Tel: +886 (3) 5644185  Fax: +886 (3) 5644170                       Add: Altenhofstrasse 80, D-66386 St. Ingbert, Germany 
E-mail:

 

[email protected]  Http://www.radiantek.com.tw 

Add: 1F, No.3, Industrial East 9

th

 Road, Science-Based Industrial Park, 

HsinChu, Taiwan 300. 

Date of issue: 2020-06-15 
Ref No.: 062020 

Reviews: