background image

 

Symbol Descriptions 

 

The CE mark and 

Notified Body 

Registration Numbers, 

the requirement of 

Annex II from Medical 

Device Directive 

93/42/EEC are met.

 

 

Please do not dispose of 

the product in the 

household waste at the 

end of its useful life. 

Disposal can take place at 

appropriate collection 

points provided in your 

country. 

Authorized 

representative in 

the European 

community 

 

Caution 

 

Please read the 

instructions for use 

Paper Recycling 

 

Manufacturer 

 

BF type applied part 

 IP22

 

Classification for 

water ingress and 

particulate matter. 

 

To protect the environment, dispose of empty batteries at your retail store or at appropriate collection sites 

according to national or local regulations. 

 

 

Manufacturer’s declaration – electromagnetic immunity 

The 

TH0xyz series

  is intended for use in the electromagnetic environment (for home healthcare) 

specified below. The customer or the user of the 

TH0xyz series

  should assure that it is used in such an 

environment. 

Immunity test 

IEC 60601 test 
level
 

Compliance 

level 

Electromagnetic environment – 

guidance 

(for home healthcare environment) 

Electrostatic 
discharge(ESD) 
IEC 61000-4-2 

Contact:

±

8 kV 

Air

±

2 kV,

±

kV,

±

8 kV,

±

15 

kV

 

Contact:

±

8 kV   

Air

±

2 kV,

±

kV,

±

8 kV,

±

15 

kV

 

Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30% 

Power 
frequency(50, 60 
Hz) magnetic field 
IEC 61000-4-8 

30 A/m 

50 Hz or 60 Hz

 

30 A/m 

50 Hz and 60 Hz

The

TH0xyz series

 power  frequency 

magnetic fields should be at levels 
characteristic of a typical location in a 
typical home healthcare environment. 

 

Manufacturer’s declaration – electromagnetic immunity 

The 

TH0xyz series

  is intended for use in the electromagnetic environment (for home healthcare) 

specified below. The customer or the user of the 

TH0xyz series

  should assure that it is used in such an 

environment. 

Immunity test 

IEC 60601 test 

level 

Complianc

e level 

Electromagnetic environment – guidance 

(for home healthcare environment) 

Radiated RF   
IEC 61000-4-3 

10 V/m 
80 MHz – 2,7 
GHz  
80 % AM at 1 
kHz

 

10 V/m 
80 MHz – 
2,7 GHz   
80 % AM at 
1 kHz 

 

Recommended separation distance: 
d = 1,2 

√P

 

d = 1,2 

√P

    80MHz to 800 MHz 

d = 2,3 

√P

    800MHz to 2,7 GHz   

 
Where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and is the recommended separation 
distance in metres (m). 
 
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey, 

a

 

should be less than the compliance level in each 
frequency range.

 

Interference may occur in the vicinity 
of equipment marked with the 
following symbol: 

NOTE1    At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE2    These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people. 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and 
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted 
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an 
electromagnetic site survey should be considered. If the measured field strength in the location in which 
the 

TH0xyz series

 is used exceeds the applicable RF compliance level above, the 

TH0xyz series

 

should be observed to verify normal operation. If abnormal performance is observed, additional measures 
may be necessary, such as re-orienting or relocating the 

TH0xyz series

 

Manufacturer’s declaration-electromagnetic emissions

The 

TH0xyz series

  is intended for use in the electromagnetic environment (for home healthcare) 

specified below.The customer or the user of the 

TH0xyz series

  should assure that it is used in such an 

environment.

 

Emissions test

 

Compliance

 

Electromagnetic environment – guidance 

(for home healthcare environment)

 

RF emissions 
CISPR 11

 

Group 1

 

The 

TH0xyz series

  uses RF energy only for its internal 

function. Therefore, its RF emissions are very low and are not 
likely to cause any interference in nearby electronic 
equipment. 

 

RF emissions 
CISPR 11

 

Class B

 

The 

TH0xyz series

  is suitable for use in all establishments, 

including domestic establishments and those directly 
connected to the public low-voltage power supply network 
that supplies buildings used for domestic purposes.