17
Table 6
recommended separation distances between
portable and mobile rF communications equipment and the Kyrobak
The Kyrobak is intended f or us e in an electromagnetic environment in which radiated rF disturbances are
Controlled. The customer or the user of t Kyrobak can help prevent electromagnetic interference by maintaining
a minimum distance between port able and m obi le rF communications equipment (transmitters)and the Kyrobak as recommended
below, according t o t he maximum out put power of t he communications equipment
rated maximum
output power of
transmitter
Watts [W]
Separation distance according to frequency of transmitter
Meters [m]
150kHz to 80MHz
d = 1.17√P
80MHz to 800MHz
d = 1.17√P
800MHz to 2.5GHz
d= 2.34√P
0.01
0.12
0.12
0.24
0.1
0.37
0.37
0.74
1
1.17
1.17
2.34
10
3.7
3.7
7.4
100
11.7
11.7
23.4
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
nOTe 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.