User Manual 2460 Series
Installation Instructions 2-4
Racal Instruments
2003
The VXI
plug&play
Driver and
Product Manuals
CD part number 980833-003 supplied with this unit contains the
2460 series driver. To install the VXI
plug&play
driver, the
computer must be running WIN 95 or WIN NT.
NOTE: This CD also contains the 2460 Series manual part
number 980833 and the 2461C Universal Counter Timer manual
980833-002.
Checking To See
That The Module
is Operational
Removing the
Module From The
Mainframe
Power-up the chassis/mainframe. Ensure that the SYSFAIL LED
on the module’s front panel illuminates and then extinguishes
after a few seconds. This indicates that the module has passed
its power-up self test, and is now ready for use.
1. Initialize the system as appropriate (eg. for a PCI MXI-2
system) run Resman
2. Run the Self-Test from the VXI
plug&play
soft front panel
and check that a PASS is indicated.
To remove the module from the VXI mainframe, power-down the
mainframe, and release the captive screws that secure the
module to the mainframe. Use the plastic levers provided on the
top and bottom edges to eject the module from the mainframe.
Pull the module along the guides provided and out of the
mainframe.
Recommended
Controllers
To utilize the full capability of the module, the slot 0 controller
should have A32 addressing capability. A controller with only
A16/A24 capability may be used but the speed advantage gained
with shared memory access will not be realized. Note: the
module has no A24 capability.