19
support.R82.org
Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to
maintain a device in accordance with the instructions may invalidate the device’s warranty. Furthermore,
failure to maintain a device may compromise the clinical condition or safety of users and/or their
caregivers.
1st Day
• Read the User Guide carefully before
use and store it in a safe place for future
reference
• Store the enclosed tools on/at the product
Daily
• Use a dry cloth to clean the product
• Visual inspection. Repair or replace any
damaged or worn-out parts
• Ensure that all fasteners are in place and
fitted correctly
•
Check all fittings and buckles for signs of
worn out parts
• Be aware of any MAX mark indications
before adjustment
Weekly
• Check that all castors are moving freely
and that all wheel locks/brakes are easily
activated
• Wipe off any residue and dirt from the
product using a cloth with warm water and a
mild detergent/soap and dry before use
• Check that the angle mechanism functions
correctly and that the safety lock is working
correctly
• Disinfect the product. please check the
disinfection section in this User Guide for
more information
Monthly
• All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
• Oil the swing-away parts. We recommend
you use a professional lubricant system
Annually
• Inspect the frame for cracks or signs for
worn-out parts and carry out annual service.
Never use a product which appears to be
substandard or faulty
Warewashing machine
This product can be washed at:
• 60°C in 10 minutes
• 85°C in 3 minutes
...using a mild detergent in a washing machine
designed for medical devices. Use the machine’s
drying function to dry the product. Clean the
accessories separately.
Before washing
Remove the following items and clean them
separately:
• Accessories
• Cushioning
• Spoke protectors
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to
wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before disinfection.
Service interval
A detailed inspection of the product must be
made every 12 months (every 6 months for heavy
users) and every time the product is reissued for
use. The inspection should be carried out by a
technically competent person who has been
trained in the use of the product.
The lifetime of this product in normal use is 5
years if all maintenance and servicing is carried
out in accordance with the manufacturer’s
instructions and demonstrably recorded.
EN
Summary of Contents for Rabbit Series
Page 1: ...support R82 org 2020 02 rev 8 EN 12182 0 30 70 kg 154 3 lb Rabbit User Guide...
Page 4: ...4 support R82 org 2 4 5 3 1...
Page 5: ...5 support R82 org 1 2 4 5 3 STORAGE...
Page 6: ...6 support R82 org 1 2 3 5 6 4 MAX...
Page 7: ...7 support R82 org 5 mm 5 mm 2 3 5 4 1 6...
Page 8: ...8 support R82 org 5 mm 2 5 Nm 5 mm 2 5 Nm 8 9 11 10 7 12 x2...
Page 9: ...9 support R82 org 3 2 5 4 1 6...
Page 54: ...54 support R82 org R82 R82 a 41 R82 9996097649 r82 org support EN 1021 1 EN 1021 2 RU...
Page 55: ...55 support R82 org 1 60 C 10 85 C 3 70 IPA 12 6 5 RU...
Page 63: ...63 support R82 org SN R82 41 C R82 9996097649 r82 org support EN 1021 1 EN 1021 2 ISO7176 16...
Page 64: ...64 support R82 org SN SN 60 C 10 85 C 3 70 IPA 5...
Page 66: ...66 support R82 org JP R82 41 C R82 9996097649 r82 org support EN 1021 1 EN 1021 2 ISO 7176 16...
Page 67: ...67 support R82 org JP 60 10 85 3 70 12 6 5...
Page 69: ...69 support R82 org GR R82 R82 41 C R82 9996097649 r82 org support exspose EN 1021 1 EN 1021 2...
Page 70: ...70 support R82 org GR 1 60 C 10 85 C 3 IPA 70 12 6 5...
Page 71: ...support R82 org...