R-Biopharm Lumitester PD-20 Brief Informations Download Page 2

8

Equipment / Software / Accessories

6

7

9

8

10

Put the swab back into the tube and 
push it downwards until the holder is 
completely inserted into the tube. 

Shake the LuciPac Pen repeatedly in 
direction of arrow (horizontal) so that  
the liquid in the capsule falls into the 
reaction tube and dissolves the regent. 

Insert the LuciPac Pen into the measuring 
chamber of the Lumitester PD-20 and 
close the chamber cover. 

Press ENTER to start the measurement. 
The result will be displayed after a  
countdown from 10 to 0.  
Keep instrument steady in an upright 
position during the measurement. 

Remove LuciPac Pen from the chamber 
after measurement. 

1

2

4

3

5

Take LuciPac Pen tubes out of the  
refrigerator 20 min prior to use and  
bring them to room temperature. 
Do not use cold Lucipac Pen test tubes. 

Turn on Lumitester PD-20 by pressing 
POWER button. 

Remove swab stick (orange holder)  
from the test tube. 

Moisten swab with tap water.  

Swab the surface which you want to test. 

20  

 min.