10
Put the swab back into the tube and
push it downwards until the holder is
completely inserted into the tube.
Shake the LuciPac Pen repeatedly in
direction of arrow (horizontal) so that
the liquid in the capsule falls into the
reaction tube and dissolves the regent.
Insert the LuciPac Pen into the measuring
chamber of the Lumitester PD-20 and
close the chamber cover.
Press ENTER to start the measurement.
The result will be displayed after a
countdown from 10 to 0.
Keep instrument steady in an upright
position during the measurement.
Remove LuciPac Pen from the chamber
after measurement.
1
2
4
3
5
Take LuciPac Pen tubes out of the
refrigerator 20 min prior to use and
bring them to room temperature.
Do not use cold Lucipac Pen test tubes.
Turn on Lumitester PD-20 by pressing
POWER button.
Remove swab stick (orange holder)
from the test tube.
Moisten swab with tap water.
Swab the surface which you want to test.
20
min.