qutools quED Manual Download Page 3

1.1. QUICKSTART MANUAL

1.1.2 Starting up

waveplate and

magnetic interlock

longpass filter

(pump block)

slot for alignment plate

pump and source

beam folding

mirror

polarizer in

rotation mount

longpass filter

fiber coupler

polarization maintaining

longpass filter

fiber

ˆ Be careful NOT to move the adjusting knobs of the mirror mounts and the ber

couplers when unpacking the breadboard!

ˆ Plug the bers into the quCR control unit.
ˆ Connect the Laser unit to the quCR control unit using the included SUB-D9 cable.
ˆ Make sure the half waveplate is unpacked and inserted into the beam path properly.

You can access the waveplate by removing the small round cover in the white source

enclosure. Please note that removing the cover will activate the safety interlock and

shut o the laser, if it was turned on.

ˆ Make sure the polarizers are both inserted securely in the two beam paths of the

optical setup. If you have the motorized version, please note that the third motor

position in the left beam path of the quED can only be used by a horizontally

mounted motor! Vertically mounted motors can only be xed in the other two

positions.

ˆ Probably the system will have to be slightly readjusted. For this purpose, set both

polarizers to

45

and the integration time to 0.1 s. Keep the laser diode covered

under the white housing and turn on the laser. Set the operating current specied

in the inspection data sheet.

ˆ A small coincidence rate should already be visible. Otherwise, you have to refer

to the complete alignment procedure in the operation manual. Make sure, the

coincidence rate is not comprised of accidental coincident events only. A rate of a

few hundred per second is usually ne and enough to start with.

www.qutools.com

quED Manual V2.0 Page 4

Reviews: