Quorum Q150T Fitting Instructions Download Page 1

 

Page 1 of 2 

 

 

Fitting instructions for sputter insert Q150T 

 
Check the machine is not operating. 

 

Pull the Silver Knob on the top cover to release the latch.  

Lift the top cover and lay back against its stop. 

 

Turn the two silver securing catch knobs to release the installed insert and remove 

the insert. 

 

Note! If the insert is a carbon rod type you may have to remove the rods first. To do 

this check that the silver securing catch knobs are restraining the insert and open the 

lid assembly and lay back against the top cover assembly. Then, using a 2.5mm hex 

key, release the rod clamping screws and withdraw the rods. Close the lid assembly 

and release the insert by turning the silver securing catch knobs. Now remove the 

insert. 

 
Prepare the sputter insert for installation by removing from any packaging, plastic 

bags etc. Check the ‘O’ ring in the outer ring of the lid assembly for damage or debris 

and clean or replace as required. Then place the insert in position in the outer ring of 

the lid assembly carefully locating the M3 screw head in the rim of the insert in the 

slot to the right of the outer ring. Lock the insert in position by turning the silver 

securing catch knobs until their flanges overlap the insert. 

  

Plug the connector on the end of the sputter insert lead in to the socket on the hinge 

assembly. 

 

To fit a target to the sputter head first open the top cover 

assembly and lay back against its stop. 

 

Wearing latex gloves, unscrew the target clamp ring and hold a 
new target in place on the holder with one finger through the 

clamp ring. Screw the clamp ring firmly back in place. The target 

should be held concentric to the holder. 

 

 

 

Close the top cover assembly and press firmly down to secure the 