Combined RF Upconverter/Downconverter
IEEE 488.2 Command Reference
7.12 *STB?
The
*STB?
(STatus Byte) query returns bit 0 to 5 and bit 7 of the QM1004A-2-18 Combined RF Upcon-
verter/Downconverter status byte and returns the Master Summary Status (MSS) as bit 6. The MSS is
inclusive OR of the bitwise combination (excluding bit 6) of the Status Byte and the Service Request En-
able registers. The format of the return is an integer between 0 and 255. Table 4.14 shows the contents
of this register.
Table 4.14:
*STB? mapping
Bit
Base 2
Meaning
0
1
Not used
1
2
Device Dependent
0 - No device status condition has occurred
1- A device status condition has occurred
2
4
Error/Event Queue
0 = Queue empty
1 = Queue not empty
3
8
Questionable Status Summary
0 - No QUEStionable status conditions have occurred
1 - A QUEStionable status condition has occurred
4
16
Message Available
0 - no output messages are ready
1 - an output message is ready
5
32
Event Status Bit
0 - no event status has occurred
1 - an event status condition has occurred
6
64
Master Summary Status
0 - Combined RF Upconverter/Downconverter not re-
questing service
1 - there is at least one reason for requesting service
7
128
Operation Status Summary
0 = No OPERation status conditions have occurred
1 = An OPERation status condition has occurred
Syntax
*
STB
?
Quonset Microwave
Revision 1.4.3
83