Quincy lab 140-1 Series Operating Manual Download Page 3

PAGE 3

Push the illuminated power button. Rotate the thermostat knob clockwise to approximately the number 
6 on the dial. The heat cycle light will illuminate to indicate that the heating element is energized.

Observe the thermometer.  When the desired temperature is reached, slowly turn the thermostat knob 
counter clockwise until the heat cycle indicator turns off. The temperature will continue to rise slightly 
due to latent heat from the heating element.

Allow unit to cycle empty for a minimum of 20 minutes to allow the interior chamber to reach a thermal 
equilibrium. Check the thermometer to see if any minor adjustments to the thermostat should be made. 
With any subsequent adjustment allow time for chamber to achieve a steady state before determining if 
the desired set temperature has been reached. To keep from resetting a frequently used set  
temperature with each use, leave the thermostat knob at its set point and use the power button to turn 
the unit on or off.

Setting the unit's temperature with an empty chamber helps to establish a useful correlation between 
the unit's performance (heating capacity) and proper processing of various chamber loads and 
densities (load-effect). See Performance and Chamber Loading & Important Operational Notes.

FIG. 7

FIG. 5

FIG. 6

(cont.)

Understanding the unit's light pressure thermal convection and "load-effect" are necessary to 
optimizing unit performance. Article or media processing times and/or uniformity are largely dependent 
on load density and positioning. When processing various loads it's important to remember that the 
thermostat senses the temperature at the lower part of the chamber where the heat is generated and 
the corresponding reading of a temperature setting is taken from the top of the chamber with the 
installed thermometer. Since loads sit between these two points, load variations such as quantity, 
arrangement, density and their relative thermal properties can temporarily or permanently affect 
temperature readings. Here are important guidelines to chamber loading and processing:

Load the incubator so that air circulation within the chamber is not impaired. Leave a space between 

articles on a shelf. Stagger articles from those on lower shelves in a "V" formation. 

(FIG 5).

Use of large solid trays or foil on lower shelves severely limits heat to shelves and articles placed 

above. 

(FIG 6)

 Since not enough heat rises within the chamber, thermometer readings give false 

indication that the temperature setting is too low. Higher temperature adjustments made as a result 
of these readings could overheat lower placed articles or media.

General Operation

Performance and Chamber Loading

Reviews: