PAGE 1
Issue Rev A 0309
Quincy Lab Inc. 1925 N. Leamington Ave. Chicago, Illinois 60639 1-800-482-HEAT (4328)
Copyright Quincy Lab Inc. 2009
Model Series 140-1 & 180-1 (230 V )
General Purpose Incubators
OPERATING MANUAL
Quincy Lab, Inc.
HEAT/CYC
LE
POWER
Model 12-140 Incubator
Interior Dimensions
INCHES W x H x D
(CM) W x H x D
Exterior Dimensions
INCHES W x H x D
(CM) W x H x D
12x10x10
18x16x12
18x16x12
12x10x10
13x15x11
19x21x13
13x15x11
19x21x11
Weight (lbs)
19 lbs
33 lbs
19 lbs
33 lbs
Standard Electrical: 230 volts, 50/60 Hz
31x25x25
46x41x30
46x41x30
31x25x25
33x38x28
48x53x33
33x38x28
48x53x33
SPECIFICATIONS
MODEL
10-140-1
MODEL
12-140-1
MODEL
10-180-1
MODEL
12-180-1
Temperature Range
A 2
o
C to 62
o
C
Common Unit Construction
Exterior
: Powder-Coated Steel
Interior:
Aluminum
Insulation:
Fiberglass
Door:
140-1: Acrylic, 180-1: Steel Insulated
Thermo-control:
Bi-Metal
Heater:
Resistive-Tubular Incoloy
WATTS
120
235
270
385
HEAT/CYCLE
POWER
Model 10-180
Incubator
A 3
o
C to 94
o
C
Cubic Foot Capacity
.7 ft
3
2.0 ft
3
.7 ft
3
2.0 ft
3
Model 10-180-1
Model 12-140-1
Common Unit Specifications
Operating Environment
:
Indoor use, altitude to 6,500 ft. (2,000m) Installation Category II,
Pollution Degree 2, ambient temperature 10
o
C/50
o
F to 35
o
C/95
o
F,
80% RH maximum.
Storage Temperature:
-10
o
C/14
o
F to 70
o
C/158
o
F, 70% RH maximum.
Approvals:
Underwriter's Laboratory Listed, Laboratory Equipment, C/UL
United States/Canadian. E212550 (115VAC models only)
Compliance:
UL Standard 61010-1, IEC 61010-1, 2nd Edition.