5. TROUBLESHOOTING TIPS

Failure of HERS testing may be caused by a number of factors. 

Here is a list of troubleshooting tips to assist in detecting some possible causes of HERS

failure that pertain to this testing equipment, and apparatus.

•  If it is ducted whole house fan, examine ductwork for improper installation including too 

sharp of a bend, too many bends, excessive length or any obstructions on exhaust side of 

fan

•  Verify that the WFH-PFH hood does not have any rips or tears on the fabric

•  Verify the airflow measurement taps in the hood are not in path of the airflow

•  Verify all tube connections are properly connected for airflow measurement apparatus and 

hood adapter system.

•  Ensure proper connection of the WHF-PFH flow hood to testing fan apparatus has a tight 

seal at the collar to fan connection. Leakages at the collar will produce inaccurate results, 

and may not allow the hood to reach 0 Pa required for HERS testing.

•  Ensure proper connection of the WHF-PFH flow hood to ceiling has a tight seal at all four 

corners of the hood. Leakages at the hood to ceiling connection may produce inaccurate 

results, and may not allow the hood to reach 0 Pa required for HERS testing.

•  Be sure to use calibrated airflow measurement apparatus when performing HERS testing. 

Refer to manufacturer manuals for keeping calibrations current.

Summary of Contents for WHF-PFH

Page 1: ...OPERATION MANUAL HERS POWERED FLOW HOOD MODEL WHF PFH MAKERS OF THE QUIETCOOL ADVANCED WHOLE HOUSE FAN...

Page 2: ...ce personnel should service the system The user should not attempt to service the product 6 Maneuvering the Unit When lifting stand or moving the unit around the house be very careful not to drag it a...

Page 3: ...Velcro iv Coroplast rectangular cutout v Soft plastic tubing with 4 barbed tee fittings pressure taps vi Fiberglass poles QTY 4 Bottom square Frame 27 5 x 27 5 Top Octagonal Frame Telescoping Legs i 4...

Page 4: ...Do not route cord under furniture or appliances Arrange cord away from traffic area and where it will not be tripped over Do not place this equipment or power cord in water or other liquid Use only t...

Page 5: ...needs to be between 4 5 from ceiling ii It is recommended to adjust height of legs prior to adding top octagonal frame c FOR RETROTEC FANS i Slide Retrotec fan in between the 4 legs See pictures below...

Page 6: ...completely or test may yield inaccurate results 4 Ensure hood is flush by lightly pushing up on bottom of hood a If you see the top of hood move readjust poles by going higher b Check again and repeat...

Page 7: ...Gauge 1 Take the pressure tap coming out the bottom of the flow hood and connect it to the input tap to read pressure 2 See page 9 for complete connection details Complete WHF PFH with TEC Model 3 Fa...

Page 8: ...procedures 1 Open the window s that are typically opened during WHF operation 2 Turn on all WHFs required to meet the dwelling unit WHF airflow rate Adjust multiple or variable speed WHFs to operate...

Page 9: ...he channel A pressure channel reference tap open to read pressure inside the dwelling 5 On TEC Model 3 fan connect the pressure tap from the blower door fan to the channel B flow channel input tap 6 O...

Page 10: ...e airflow Verify all tube connections are properly connected for airflow measurement apparatus and hood adapter system Ensure proper connection of the WHF PFH flow hood to testing fan apparatus has a...

Page 11: ...ow rate listed on CF 1R report Title 24 reports that have modeled medium speeds for 2019 code only need to be tested at medium speed If the measured flow rate is failing too low try running WHF at a h...

Page 12: ...12 Rev 6 11 20 MAKERS OF THE QUIETCOOL ADVANCED WHOLE HOUSE FAN...

Reviews: