background image

Sofia and Virena Router Connection

1.

 Power OFF Sofia. 

2.

 

Connect the LAN Cable: 

Connect one end of the Ethernet cable to the “LAN 0” port of the router and the other end to a LAN port on the back panel of Sofia (Figure 2).

3.

 Power ON Sofia. 

1253802  (03/14)

Set Network Configuration on Sofia

1.

 Go to Main Menu     Supervisor Menu (log in as Supervisor)     Settings     General Settings     

Set Network 

Configuration.

2.

 Scroll to 

“Set IP Address:”

 and ensure 

Manual

 is checked, see Figure 4. Scroll to 

“Network Address”

 and 

enter IP network address to 192.168.0.5. Verify that all configurations on your device including: Subnet Mask, 

Gateway and Port Number match what is shown in Figure 4. If it does not match, please enter the correct 

numbers. Select OK.

3.

 Screen will display "Please wait." Once this message disappears you will be back on the General Analyzer 

Settings menu. Select the "Back" button twice to get back to the Supervisor Menu     Supervisor Menu     

Statistics     

View Network Settings.

 Ensure that you have a valid IP address.

 

Note:

 The IP address will automatically be entered if Sofia is already connected to the Virena Router. 

Surveillance Mode Set Up

1.

 Go to Main Menu     Supervisor Menu     Settings     

Surveillance Settings.

  Note:

 If Sofia does not show 

Surveillance Settings

 as an option, Sofia will require an upgrade to firmware 

1.3.1 or above. For further assistance, please contact Quidel Technical Support at 800.874.1517 (in the U.S.), 

858.552.1100 (outside the U.S.) or e-mail [email protected].

2.

 In 

“Set SURV Parameters”

 screen (Figure 5) and select the following:

  Age Required:  

 Yes  

  Transmit Patient ID:  

✔ 

Off 

  Auto Send:  

 On

  Queue Results for SURV:  

 On

3.

 When completed, press “OK”     Back     

Main Menu.

Virena Router Connection Verification

1.

 Go to Main Menu     Review Data     Send Results     Send Results to SURV Receiver     

Test SURV 

Connection.

 

2.

 If connection is established, screen displays 

“The Analyzer can be connected to SURV.” 

3.

 Return to the Main Menu (Press “OK”     Back     Back); Sofia is now ready to be run in the 

Surveillance Mode.

4.

 If connection is not established, screen displays 

“The Analyzer cannot be connected to SURV.” 

5.

 Return to the Main Menu (Press “OK”     Back     Back). Review the following steps to confirm all settings and 

connections are correct:

    I.  Sofia and Virena Router Connection 

  II.  Set Network Configuration on Sofia

6.

 If unsuccessful in configuring router or network connections, contact Quidel Technical Support at

800.874.1517 (in the U.S.), 858.552.1100 (outside the U.S.) or e-mail [email protected].

Performing Tests in Surveillance Mode

1.

 Go to Main Menu     

Run Test.

2.

 Enter (User ID, Patient ID, and Order #) if required, then press “

Start Test.”

3.

 

Patient Demographics

 screen is displayed (Figure 6).

 

Note:

 Fields #1-4 allow for additional demographics data not currently defined in Sofia. 

4.

 Enter numeric value for 

“Age”

 (select appropriate units of Years, Months, or Weeks) by using the

right arrow key.

5.

 Press the down arrow and select 

“Patient Status” 

(select In-patient, Out-patient, or leave blank)

by using the right arrow key.

6.

 Then press

 “Start Test.”

Note:

 For additional information regarding Virena Router safety, certifications, and product disposal, refer to: www.quidel.com/immunoassays/virena. 

Figure 5

Figure 4

Figure 6

Reviews: