17
9.
Temperature Calibration
Note:
1) When the display temperature is consistent with the set temperature,
the temperature can be calibrated.
2) Long press "+" and "-" for 2s to enter the temperature calibration
interface.
For example
:
a) The product needs to be preheated to 100 ℃, so we set the
temperature to 100 ℃.
b) When the temperature is set to 100 ℃ due to the material of the
preheating plate and environmental factors, the temperature above the
preheating plate is measured to be 95 ℃ by the thermometer. Input the
value of 95 in the temperature calibration interface, and press
to
exit.
Set temperature
Display temperature