4

Zero/HEPA Filter Example

Used for particulate zero cal. 

Tubing and Vent Opening

Insert tubing into the 

impactor of the labeled “2.5” 

vent opening 

Turret

Turn the turret so “PM” 

appears on the display of 

the EVM. 

Particulate Zero Calibration

Prior to calibrating, ensure the impactors are well-greased with an applied thin-film of silicone grease. 
(Please see “Impactor Care” on page 10 of EVM series user manual.)

1. Attach the 

Zero/HEPA

 filter and tubing 

into the turret. Rotate the turret to PM.

2. Select 

Calibration

 from the 

Start

 screen 

and select 

PM

 (using up/down arrows).

3. Press 

Cal

 softkey and then press  

Start

 softkey. Once level is stabilized, 

press 

Set

 softkey.

4.   Once level has been set, you can press 

Save

 or 

Cancel

. (Please refer to  

Chapter 4 of EVM series manual for  
more information.) 

 

NOTE:  Once saved, the instrument will apply the zero calibration to all profiles. The zero calibration is 

stored in the History screen and a calibration. Verify can be performed (to ensure no drifting). 

Particulate Advanced Calibration (Gravimetric) 

 Gravimetric Calibration Overview

(1)Reset the gravimetric and volume accumulators; (2) Insert a new (sealed) gravimetric cassette;  
(3) Measure in the specific location for approximately four to eight hours; (4) Record the value displayed 
in the gravimetric accumulator field displayed in the particulate screen. (This value is used in conjunction 
with the weight of the measured dust in the cassette.); (5) Remove the gravimetric cassette and weigh 
the filtered contents. (Typically, send to a lab for results.); (6) To determine the correction factor, calculate 
the ratio between the particulate mass value and the weight from the gravimetric filter lab analysis. 
(Ratio = true weight/EVM gravimetric displayed value). (7) Enter user correction value (Setup\Particulate 
screen. Select profile field and then press Edit softkey. Add Factor using up/down arrows. Press On/OFF/
Esc key twice to return to start screen.) 

Logging Particulates

1. Before logging particulates, ensure the impactors are clean and well-greased.
2. From the start screen or any measurement screen, press           key . 
3. Press           key to end the logged session. (See “Viewing logged results” to access past sessions.)

Summary of Contents for EVM Series

Page 1: ...EVM Series Environmental Monitor QUICK START GUIDE ...

Page 2: ...een the Start screen will appear 2 Navigating to measurement screens press keys Navigating to menu options press keys To open a selected menu or to save settings press key 3 In the powered on mode press and hold key to power down the instrument Start Screen and Measurement Views You will visit the start screen frequently to access measurements setup calibration unit info and file system screens Pr...

Page 3: ...y to end your logging press key 9 Review your results either on the instrument via the File System screen or download to DMS for further analysis graphs charts saving and printing Particulate Measurements Particulate setup 1 To change particulate settings select Setup Particulate 2 Particulate screen appears and includes the following adjustable fields Particulate field enable disable pump collect...

Page 4: ...ing Particulate Advanced Calibration Gravimetric Gravimetric Calibration Overview 1 Reset the gravimetric and volume accumulators 2 Insert a new sealed gravimetric cassette 3 Measure in the specific location for approximately four to eight hours 4 Record the value displayed in the gravimetric accumulator field displayed in the particulate screen This value is used in conjunction with the weight of...

Page 5: ...t Calibration from the Start screen and select PID using up down arrows 3 Press Cal softkey and then press Zero softkey Once level is stabilized press Set softkey Typically takes 60 seconds to stabilize 4 Once level has been set you can press Save or Cancel Disconnect components and return to the start screen via On Off Esc key PID Span Calibration Typically Isobutylene is used when the VOCs are u...

Page 6: ... 50 ppm 0 1 ppm nitrogen pure air SO2 Carbon Dioxide Gas Similar to toxic gases the carbon dioxide sensor should be calibrated prior to logging Toxic and Carbon Dioxide Zero Calibration Depending on the gas of interest the toxic zero calibration is performed with the gas indicated in the table above The CO2 sensor requires a calibration gas a regulator with at least 1 0 Liter minute flow rate and ...

Page 7: ...eens by pressing from the start screen when Setup is selected 2 Press key to stop logging See Viewing logged results to access past sessions Particulate advanced calibration Gravimetric filtering Gravimetric Calibration Overview 1 Reset the gravimetric and volume accumulators 2 Insert a new sealed gravimetric cassette 3 Measure in the specific location for approximately four to eight hours 4 Notat...

Page 8: ...ate Air Flow Calibration For Airflow meter calibration an accurate flow meter such as a primary standard calibrator and tubing is required 1 Attach the primary standard calibrator tubing to the air flow outlet 2 Navigate to the Setup Calibration Flow RT rate screen 3 Press the Cal softkey and then press Start softkey 4 Adjust the flow rate using the up and down arrow keys so the primary standard c...

Page 9: ...low 3 Turn Air Velocity Probe switch On 4 Navigate to the Air Velocity measurement screen using keys Viewing Logged sessions past sessions Once you have logged data and stopped the session the data is stored under the past session summary menu Optional you can view the data in charts graphs when downloaded to DMS 1 Navigate to Start Past Session screen 2 Select the File field and press the File so...

Page 10: ...ower cord before removing or inserting sensors 1 Remove the sensor bar via the screws and remove the manifold 2 Sensor bar removal and location of sensors are displayed below When removing inserting align the socket pins to the circuit board accordingly Impactor Care and Particulates The impactor requires periodic greasing and cleaning for particulate measuring only 1 Remove the turret red housing...

Page 11: ...le into the EVM and plug the opposite end to the USB port of your computer Refer to illustration below Viewing your data In DMS once your files are downloaded via button view the data in the panel layout view or reports by clicking on appropriate buttons See an example of the panel layout page below Download ...

Page 12: ... website www tsi com for more information North America and Asia Pacific Telephone 1 800 680 1220 USA 1 651 490 2860 Outside USA Fax 1 651 490 3824 E mail technical services tsi com Europe Middle East and Africa Telephone 49 241 52303 0 Fax 49 241 52303 49 E mail tsigmbh tsi com P N 074 301 Rev E 2018 TSI Incorporated ...

Reviews: