Quest Spectrum® Laparoscopes

 

 

 

 

 

 CAUTION:  

To maintain the optical quality, make sure that the laparoscopes 
are not exposed to gamma rays. 

 Note:  

Check  image  quality  (clear  and  free  of  distortion)  and  light 
transmission through the glass fibers. 

 

Hold the light guide connector against non-glare light. Should 
the glass fibers appear as black spots on the distal end (broken 
glass fibers), it is possible that the light output is insufficient. 

 

3

 

Assembly and Disassembly  

Figure  1  shows the construction of  the  Quest  Spectrum®  Laparoscope.  The  Eyepiece can  be 
only  connected  to  the  Quest  Spectrum®  Camera.  Table  2  provides  an  overview  of  all 
laparoscopes  provided  by  Quest  Medical  Imaging  and  Table  3  provides  an  overview  of  all 
accessories.  
 

 

 

Figure 1: Construction of the Laparoscope 

 
 
(1) Objective with sapphire window 
(2) Duct 
(3) Body 
(4) Eyepiece 
(5) Color code ring (line of vision) 
(6) Adapter for light guides, removable 
(7)  Adapter  for  light  conductors, 
removable. 
(8)  Connecting  piece  for  light  guides 
(ACMI) 

 Note:  

Only  the  light  guide  connector  can  be  dissembled  from  the 
Laparoscope.  

 
 
 

Table 2: overview of all laparoscopes provided by Quest Medical Imaging. 

Article Number 

Line of vision 

Color Code 

Field of View 

Working Length  

ART-LAP-0008 

30

°

 

Red 

70

°

 

290 mm 

ART-LAP-0008S 

30

°

 

Red 

70

°

 

150 mm  

ART-LAP-0009 

0

°

 

Green 

70

°

 

290 mm  

ART-LAP-0009S 

0

°

 

Green 

70

°

 

150 mm 

 

Table 3: accessories of the Laparoscope 

Article Number 

Name 

Use 

ART-OP-0007 

Optical fiber  

Light Guide from Light Engine to Laparoscope 

 

For a safe assembly: 

  Always hold the laparoscope on the body or the eyepiece. 

  Make sure that  the adapters  of the  light-  conducting  cables  comply  with the  adapters of  the 

laparoscope (see Figure 1). 

  Make sure that glass surfaces do not interact with other instruments. 

 

Summary of Contents for ART-LAP-0008

Page 1: ...Laparoscope Instructions for Use ...

Page 2: ...Laparoscope Instructions for Use ...

Page 3: ...pport Email info quest mi com Phone 1 33 07 52 65 96 Fax 1 33 07 52 65 97 Document Identification Document QMI0024_IFU_EN_07052021_v1_1 Revision 1 1 Copyright statement 2021 Quest Medical Imaging B V All rights reserved No part of this document may be photocopied reproduced or translated in any way without the prior written consent of Quest Medical Imaging B V Every effort has been made to make th...

Page 4: ...Prior Use 4 2 3 Intraoperative Use 4 3 Assembly and Disassembly 5 3 1 Combinations 6 4 Cleaning Disinfection 6 4 1 Safe treatment 6 4 2 After every use 6 4 3 Manual cleaning disinfection 7 4 4 Machine cleaning disinfection 7 4 5 Maintenance checking 8 4 6 Packing 8 5 Sterilization 8 5 1 General Remarks 8 5 2 Sterilization method 9 5 3 Storage 9 5 4 Information on validation 9 6 Accessories spare p...

Page 5: ...arnings may result in injury to the patient or user CAUTION indicates risks to the equipment Failure to follow cautions may result in product damage Note provides special information to clarify instructions or present additional useful information 1 2 Indication for Use During the examination diagnosis and or in combination with endoscopic accessories treatment Laparoscopes serve exclusively to vi...

Page 6: ...pectrum Laparoscopes Patient or user injury and or instrument damage can result Let exclusively the respective dealer or a repair facility authorized by Quest Medical Imaging carry out repair works Use exclusively original spare parts for repair works Before sending it back Clean disinfect and sterilize the light guide cable and laparoscope thoroughly prior to returning for repair Note The guarant...

Page 7: ...disinfection and prior to every use check the laparoscope for cleanness function and damages e g insulation loose bent broken chapped rough worn broken off parts etc WARNING Do not use a damaged or defective laparoscope Contact Quest Medical Imaging 2 3 Intraoperative Use WARNING It is possible that light with a high radiation energy emerges from the viewing window and raises the temperature of th...

Page 8: ...roscope 1 Objective with sapphire window 2 Duct 3 Body 4 Eyepiece 5 Color code ring line of vision 6 Adapter for light guides removable 7 Adapter for light conductors removable 8 Connecting piece for light guides ACMI Note Only the light guide connector can be dissembled from the Laparoscope Table 2 overview of all laparoscopes provided by Quest Medical Imaging Article Number Line of vision Color ...

Page 9: ...leaning agent o vapor sterilization Use exclusively cleaning agents for complete protein dissolution WARNING Do not exceed maximum number of treatments see manual of respective product WARNING Do not use laparoscopes or accessories once their product lives are expired see manual of respective product WARNING Treat laparoscopes and accessories before the first and after every use WARNING Make sure ...

Page 10: ...cal disinfection always rinse sufficiently and thoroughly under clear running water Observe indications of disinfectant manufacturer 7 Remove stubborn dirt with a soft plastic brush Do not use abrasive cleaning agents or metal brushes 8 Clean lumens and channels using soft round plastic brushes and rinse thoroughly 9 Use fully desalinated water for final rinsing 10 Dry laparoscopes with an absorbe...

Page 11: ...essories for protein residue and dirt Clean contaminated or dirty laparoscopes and accessories again Lightly grease movable parts e g hinges and ends with serializable vapor permeable protective oil Use exclusively bio compatible paraffin oil based lubricants exception latex and rubber Assemble detachable laparoscopes and accessories After every cleaning disinfection check laparoscopes and accesso...

Page 12: ...per laminate sterilization bags Type Steriking Metal sterilization containers paper laminate sterilization bags Type Steriking Note Exclusively vapor sterilize laparoscopes marked autoclavable CAUTION Do not accelerate cooling of laparoscopes and accessories This will cause damage and reduce the product life Allow laparoscopes and accessories to cool down to room temperature 23 2 C Open packages a...

Page 13: ...rse endoscopes and accessories in a non corrosive non fixing disinfection bath that is changed daily Leave endoscopes and accessories in disinfection bath for a maximum of 8 hours Before machine cleaning and disinfection rinse endoscopes and accessories thoroughly under clear running water 8 1 2 Dry Disposal Machine clean endoscopes and accessories For the disposal mind the following Dispose of th...

Page 14: ...m Laparoscope bear the CE mark Quest Medical Imaging B V herewith declares that a We have verified the mutual compatibility of the devices of the Quest Spectrum Laparoscope in accordance with the manufacturers instructions and have carried out his operations in accordance with these instructions b We have packaged the Quest Spectrum Laparoscope and supplied relevant information to users incorporat...

Reviews: