Rack-Mount RDMS
TM
Telemetry Receiver
71
Quasonix, Inc.
Figure 127: Mod Scaling Menu, Hold Threshold Selection
A dialog screen displays prompting the operator to enter a new modulation scale hold threshold using the rack’s
front panel numeric keypad. The Mod Scale Hold Threshold Entry screen is shown in Figure 128. The allowable
range is -20.000 to +30.000. When the new value is entered, press the Enter key on the keypad.
Figure 128: Modulation Scaling Hold Threshold Entry Screen
4.1.5.19 Encoding Menu
Access the Encoding menu from the Advanced Menu. Select ‘Encoding Menu’, shown in Figure 129, then press the
Enter key on the front panel keypad.