![Quartzlock A5000 User Handbook Download Page 4](http://html1.mh-extra.com/html/quartzlock/a5000/a5000_user-handbook_3535286004.webp)
Model A5000 Operation Manual
A5000 Manual A5.docx 19 August 2008
Page 4 of 5
Voltage, Frequency and Power Characteristics
Voltage 220-240V AC
Frequency 40-50Hz
Power characteristics 500mA Max
Environmental Conditions
Temperature
Operating (ambient)
-10°C to +55°C (-65 to +65 op)
Storage
-40°C to +85°C
Magnetic Field
Sensitivity
≤
2x10
-11
/ Gauss
Atmospheric Pressure
-60m to 4000m
<1x10
-13
/ mbar
Replaceable Fusing Characteristics
800mA time-lag HBC
Cleaning Instructions
To ensure long and trouble operation, keep the unit free from dust and use care with liquids around the unit.
Be careful not to spill liquids onto the unit. If the unit does get wet, turn the power off immediately and let the unit dry completely
before turning it on again.
Clean with a damp (with water) cloth.
Never spray cleaner directly onto the unit or let liquid run into any part of it. Never use harsh or caustic products to clean the unit.
Distribution Amplifier A5000
Operating Procedure
Introduction
The Quartzlock model A5000 is a distribution amplifier designed for standard frequency distribution between 1 MHz and 100 MHz.
Input
One input is provided. This has input impedance of 50 . The input level should be between 0 and +13 dBm.
Outputs
Up to six outputs are provided. These have source impedance of 50 , and will provide between 0 and +13 dBm into a 50 load.