Model
A5
-12 Operation & Service Manual
A5-12 Manual A5 21 January 2005
Page 11
1. Remove the lid.
2. Connect desired input between 0 and 13 dBm.
3. Connect a spectrum analyser, oscilloscope with 50
Ω
input impedance, or RF
power meter to any output.
4. Adjust output level to 13dBm (1V RMS) using the pre-set (VR1) on the main
board.
5. Replace the lid.
4.2.9
Note:
All the amplifiers are class A linear amplifiers. The gain may be set to provide any
output level up to clipping if desired.