Model A3-R Quartz Frequency Standard 

 

Introduction 

The Model 3 quartz frequency standard is based upon a new design of oven – controlled 
crystal oscillator and will give excellent service over long time periods. The unit is fed from an 
internal, switched – mode power supply working at 50/60 Hz and supplied from a mains source of 
between 110 V and 240 V (r.m.s.). 

 
Front Panel 

The front panel is fitted with a green “power on” light – emitting diode indicator on the right – hand 
side. 

 
Back Panel 
 
R.F. Output 

The back panel is fitted with a B.N.C. socket giving a 10 MHz reference output, a at level of 
approxi10 dBm (50

). This is suitable for distributing around laboratories, workbenches 

etc through a distribution amplifier. 

 
Very Long – Term Frequency Adjustments 

On leaving the factory the unit will have been adjusted against a hydrogen maser frequency 
standard, to within one part in 10

11

 
However, it is appreciated that quartz crystal oscillators will tend to drift (upwards) in 
frequency over long time periods, due to crystal ageing. Should the customer consider that the drift 
eventually renders the unit in need of correction, the following should be carried out: 
 
EITHER Return the unit to Quartzlock for re – adjustment against the hydrogen maser there 
 
OR If a high accuracy frequency standard is locally available, remove the six lid – 
retaining screws and locate the blue, 10 kR, 20 turn potentiometer, next to the OCXO unit.  
Very carefully adjust this potentiometer until the frequency is within acceptable limits. A 
comparison frequency standard, whose accuracy is better than one part in 10E-12, is recommended 
for this adjustment. 

Internal Switched-Mode Power Supply 

 

WARNING:  Risk of electrical shock or burns 

Capacitors store hazardous energy 
Do not touch the power supply until after 5 minutes after disconnecting the mains 

 
Pin Connections 
 

1. 

AC(L) 

Input terminal live 

2. 

AC(N) 

Input terminal neutral 

3. 

FG 

Field Ground (Protective Earth) 

4. 

+ Vout 

Output terminal VCC 

5. 

- Vout 

Output terminal GND 

6. 

Vadj. 

Output voltage adjustment potentiometer 

 