a) Section «MODE»
The section allows you to choose either «MEASURE» profile or evaluative
«SEARCH» profile.
Move the cursor by pressing « ». Switch on the mode with the button «SELECT»
and it is accompanied by the setting of a tick « » in front of the desired line.
Press of «RETURN» returns you to the main «screen» of the menu.
b) Section «THRESHOLD»
The section «THRESHOLD» allows you to set the value of dose rate and flux
density above which the alarm gets activated.
The threshold alarm should be switched on if the call signal or vibra signal re-
quired only after exceeding a specified value of measurement. The user sets the alarm
threshold within the range of the threshold value from 0,05 to 1,20 µSv/h and from
5 to 120 1/cm
2
*min personally.
Note-The factory settings are:
dose threshold – « 0,30 µSv/h»,
flux density threshold - «10 1/cm
2
* min»
→
1 2 3
4 5 6
7 8 9
►
√
MEASURE
SEARCH
●
CHANGE RETURN
●
●
▼
►
GAMMA
BETA
●
SELECT RETURN
●
●
▼
MODE
►
THRESHOLD
BACKGROUND
SETTINGS
CALIBRATION
●
SELECT RETURN
●
●
▼
►
OFF
√
ON
10
1
/cm
2
*min
●
CHANGE RETURN
●
●
▼
OFF
ON
√
10 1
/cm
2
*min
●
+ 5
SAVE
●
●
- 5
OFF
ON
√
50 1
/cm
2
*min
●
+ 10
SAVE
●
●
- 10
OFF
ON
√
120 1
/cm
2
*min
●
RESET
SAVE
●
●
- 10
OFF
ON
√
0 1
/cm
2
*min
●
+ 5
●
МАХ
SAVE
●
OFF
√
ON
►
10
1
/cm
2
*min
●
CHANGE
RETURN
●
●
▼
►
MODE
THRESHOLD
BACKGROUND
SETTINGS
CALIBRATION
●
SELECT
RETURN
●
● ▼
GAMMA
►
BETA
●
SELECT RETURN
●
●
▼
14