5. Re-connect the air tubes.
The Control Unit is not working
1. Check the power cord and the power voltage.
2. Check the fuse
1. Use a power regulator
2. Replace with a new fuse
Some air cells have abnormal low air pressure
while the air pressure for other air cells is
normal
1. Is the connection between air cells and the manifold
kinked?
2. Is there any air leakage from the air cells?
1..Adjust the connection between cells
and manifold
2. Replace with a new air cell.