10. Position the collimator at 40" SID.
11. Pull the cassette film tray handle out so it extends out of the front opening in
table.
12. Turn on the collimator light.
13. Check for proper alignment of the image receptor centering light (emitted by
the collimator) with the notch in the cassette film tray handle. If misaligned,
refer to the collimator manual for alignment procedures.
TABLETOP ASSEMBLY/INSTALLATION
CAUTION! Be careful not to damage the Obstruc-
tion Sensors when installing the Tabletop Assem-
bly.
The following procedure describes how to install the Tabletop Assembly in the
Tabletop Frame.
1. Unpack the Tabletop Assembly from the carton.
2. At one end of the Tabletop Assembly, remove three (3) 1/4-20 x 1" low profile
socket head screws securing End Trim Plate to Tabletop Assembly.
3. Remove two (2) Table Top End Stop Bumpers from Tabletop Frame (see Fig-
END STOP
BUMPERS
END TRIM
PLATE
Figure 6. Table Top End Trim Plate and End Stop Bumpers
TABLE TOP
ASSEMBLY
END
TABLETOP
SUPPORT
Summary of Contents for QT-740
Page 4: ...Revision History ii Models QT 740 QT 750 Service Manual THIS PAGE INTENTIONALLY LEFT BLANK...
Page 7: ...Chapter 1 1 SPECIFICATIONS...
Page 8: ...2...
Page 13: ...Chapter 7 2 ASSEMBLY INSTALLATION...
Page 14: ...8...
Page 29: ...Chapter 23 3 THEORY OF OPERATION...
Page 30: ...24...
Page 39: ...Chapter 33 4 SERVICE INSTRUCTIONS...
Page 40: ...34...
Page 57: ...Chapter 4 Service Instructions 51 Models QT 740 QT 750 Service Manual...
Page 58: ...Chapter 51 5 DIAGRAMS...
Page 59: ...52...