Quantum QT-740 Operator'S Manual Download Page 15

 

Chapter 2  General Information

Models QT-740 & QT-750 - Operator’s Manual

9

OVERVIEW

This manual provides information for installing, operating, and maintaining Quantum 

Medical Imaging’s Fixed (Model QT-740) and Elevating (Model QT-750) Radiographic 

Tables (hereinafter referred to as the radiographic table).  It is imperative that all 

safety procedures described in this manual be strictly adhered to in order to ensure 

the safety of both patient and user.
The key features of the Model QT-740 and QT-750 radiographic tables are as follows:

Large Tabletop area (85 inches long x 34 inches wide) provides ample 

examination platform

Four-way floating tabletop permits 36 inches of longitudinal travel and 10 

inches of lateral travel for full radiographic coverage

Foot pedal controls for table float-top motions (foot pedal inhibit switch 

deactivates foot pedals to prevent inadvertent tabletop motion)

Push button FLOAT control provided for operator convienence

Small patient-to-film plane distance provides improved radiography with 

minimal magnification

Accepts all types of image receptors (buckies, grids, and stationary grid 

cabinets)

Patient handgrips are standard accessory

AC line cord, fitted with hospital-grade plug, plugs into a 115VAC, 50/60 Hz 

receptacle (also configurable for 220 VAC, 50/60 Hz power input)

The Image Receptor Cabinet is equipped with electric locks ensuring precise 

film receptor alignment

In addition, the Model QT-750 provides:

Heavy-duty lift mechanism allows for patient loads of up to 650 lbs. (weight 

must be centered over table base)

Variable height tabletop, lowers to 21 inches and raises to 32.5 inches (11.5 

inches total vertical travel)

Foot pedal controls for table up/down motions

Collision sensors disable tabletop down motion in the event an obstruction is 

detected below tabletop

INTENDED USE

The Fixed Float-top (Model QT-740) and Elevating Float-top (Model QT-750) 

Radiographic Tables are intended for use as a patient support device during the 

performance of radiographic examinations.

NOTE 

The user should 

read this man-

ual in its en-

tirety prior to 

using this equip-

ment.  It should 

be kept in a 

location near 

the equipment 

and readily ac-

cessible to the 

those that oper-

ate it. 

Summary of Contents for QT-740

Page 1: ...Models QT 740 and QT 750 Manual Part No DC30 006 Revision A Operator s Manual Fixed Elevating Float Top Radiographic Tables...

Page 2: ...copied photocopied reproduced translated or reduced to any electronic medium or machine readable form without prior consent in writing from Quantum Medical Imaging LLC Made in U S A Quantum Medical I...

Page 3: ...Revision History Models QT 740 QT 750 Operator s Manual i REVISION DATE TYPE OF MODIFICATION A 11 01 00 Initial Release Page Number Rev Page Number Rev Page Number Rev i iv A 1 30 A...

Page 4: ...Revision History ii Models QT 740 QT 750 Operator s Manual THIS PAGE INTENTIONALLY LEFT BLANK...

Page 5: ...tents Models QT 740 QT 750 Operator s Manual iii Table of Contents Chapter 1 Safety Notices 1 Chapter 2 General Information 7 Chapter 3 Operation 11 Chapter 4 User Maintenance 21 Chapter 5 Warranty In...

Page 6: ...Table of Contents iv Models QT 740 QT 750 Operator s Manual THIS PAGE INTENTIONALLY LEFT BLANK...

Page 7: ...Chapter 1 1 SAFETY NOTICES...

Page 8: ...2...

Page 9: ...these warnings may cause serious injuries X RAY PROTECTION X rays are hazardous to both patient and operator unless established safe exposure factors and operating instructions are observed It is impo...

Page 10: ...aluation for Medical Use of X Rays and Gamma Rays of Energies up to 10 MeV as revised or replaced in the future Those authorized to operate test par ticipate in or supervise the operation of the equip...

Page 11: ...used for marking on this equipment or equipment doc umentation Earth ground Type B Equipment Protective Earth ground Attention consult accompanying documents Dangerous Voltage Enable tabletop float re...

Page 12: ...Chapter 1 Safety Notices 6 Models QT 740 QT 750 Operator s Manual THIS PAGE INTENTIONALLY LEFT BLANK...

Page 13: ...Chapter 7 2 GENERAL INFORMATION...

Page 14: ...8...

Page 15: ...ed radiography with minimal magnification Accepts all types of image receptors buckies grids and stationary grid cabinets Patient handgrips are standard accessory AC line cord fitted with hospital gra...

Page 16: ...ntains 1 Tabletop 2 Emergency Shut off Switch Model QT 750 only 3 Foot Pedal FLOAT Push Button Inhibit Switch 4 Base 5 Foot Pedal s Float Model QT 740 Float Up and Down Model QT 750 6 Float Push Butto...

Page 17: ...Chapter 3 OPERATION 11...

Page 18: ...12...

Page 19: ...of this equipment must be operated with reasonable care Manufacturer s equipment recommenda tions described in this manual must be observed WARNING To avoid damage to the table any load on table top...

Page 20: ...al positioning of the patient is enabled To operate the float top function proceed as follows WARNING When transferring a patient on or off the tabletop always make sure the tabletop is locked If nece...

Page 21: ...t pedal immediately activates tabletop locks inhibit ing transverse and longitudinal tabletop motion ELEVATING MOTION MODEL QT 750 ONLY On Model QT 750 radiographic tables the elevation function allow...

Page 22: ...e tabletop Note As it is lowered the speed at which the tabletop travels will gradually be reduced until the tabletop reaches the lowest point of travel for optimal patient comfort RAISING THE TABLETO...

Page 23: ...GE RECEPTOR CABINET LOCK The Image Receptor Cabinet rides on four linear ball bearings mounted to adjust able brackets attached to either side of the cabinet see Figure 7 These bear ings travel on the...

Page 24: ...open both grips 4 Position slide guides L brackets seated in channels to cassette size using indicators on cassette tray press center brass button to allow movement 5 Insert cassette into tray back e...

Page 25: ...from receptor using tray handle see Figure 9 3 Place thumb under lip of clamp handle Lift handle to release 4 Grasp handle and slide clamp back for inserting cassette 5 Insert cassette into tray back...

Page 26: ...Chapter 3 Operation 20 Models QT 740 QT 750 Operator s Manual THIS PAGE INTENTIONALLY LEFT BLANK...

Page 27: ...Chapter 21 4 USER MAINTENANCE...

Page 28: ...22...

Page 29: ...painted surfaces of the table Cleaning The system user is responsible for the basic cleanliness of the equipment On a regular basis the table surface should be wiped clean Painted metal sur faces sho...

Page 30: ...e cassette tray up slightly and pull the tray away from the Receptor Cabinet Poersch Trays 1 Float the table top fully back 2 Pull the cassette tray out by the handle until it stops 3 Grasp the left s...

Page 31: ...Chapter 25 5 WARRANTY INFORMATION...

Page 32: ...26...

Page 33: ...age by properly trained personnel All QMI products shall be warranted for a period of 12 months from the time of origi nal installation the date of which will be determined by a completed returned war...

Page 34: ...LY QMI Warranty is exclusive of 1 Failure of Buyer Dealer to prepare the site or provide power requirements or oper ating environmental conditions in compliance with any applicable instructions or rec...

Page 35: ...parts or assemblies All returned parts shall become the property of Quantum Medical Imaging The preceding Paragraphs set forth Buyer s Remedies and Quantum Medical Imag ing s sole liability for claim...

Page 36: ...30 Models QT 740 QT 750 Operator s Manual Chapter 5 Warranty Information THIS PAGE INTENTIONALLY LEFT BLANK...

Reviews: