 ATEX-DCT150M-W-28.02.20-00003R5 

 

HAZLOC Hazardous Area Compatibility 

 

Ensure that the gauge is suitable for use in the environment of the hazardous zone. This should be 

done by reference to the HAZLOC code of the DigiTemp which is also marked on the face of the 

instrument and stated on the certificate. 

 

Class I, II, III,  Div 2, Groups A-D, F-G, Nonincendive/ Non incendaire 

Class I, Zone 1, AEx ib IIC T4/T3 Gb, Zone 21, AEx ib IIIC T135°C /T184°C Db,  

Ex ib IIC T4/T3 Gb 

Ex ib IIIC T135°C /T184°C Db 

-35°C ≤ Ta ≤ +60°C 

Control drawing-A3-00614, Rev08 

 

 

Max. Process 

Temperature  

(Gas Atmospheres) 

Temperature 

Class 

Max. Process 

Temperature  

(Dust Atmospheres) 

Temperature 

Class 

-50°C to +105°C 

T4 

-50°C to +101°C 

T135°C 

-50°C to +150°C 

T3 

-50°C to +150°C 

T184°C 

 

 

In the event that the equipment is to be exposed to any unusual external effects or to aggressive 

substances please contact Quantum Production Ltd for further advice 

 

 

INSTALLATION 

 

Installation should only be undertaken by qualified personnel familiar with safety practices and 

standards relating to intrinsically safe electrical equipment. 

 

 

OPERATING INSTRUCTIONS 

 

The DCT150M-W samples the  temperature measured at the sensor once every minute. In practice, 

due to the slow change in the temperature of the wall and the contents of the tank being measured, 

the temperature is unlikely to change by more than a few degrees between samples. If, however, the 

sensor is removed from the tank and subjected to extremes of temperature change such as plunging 

into hot oil, for example, it is likely that the temperature seen will change by a greater degree at the 

next sample occasion, due to the fact that the temperature shown on the display is fixed between 

samples. The DCT150M is functioning perfectly correctly and merely indicating the actual temperature 

of the sensor at one minute intervals.  

As  with  any  thermometer,  it  should  be  noted  that  it  takes  time  for  the  sensor  to  reach  the 

temperature  of  the  product  measured  and  this  should  be  taken  into  account.  The  DCT150M-W  is 

wireless enabled allowing wireless transmission to compatible  remote communication devices. This 

Summary of Contents for DCT150M-W

Page 1: ... sealed polycarbonate face which is waterproof to IP67 and which eliminates surface electrochemical corrosion problems The DCT150M W is suitable for hostile environments without additional protection iNSTALLATION REQUIREMENTS Prior to installing this thermometer these instructions should be fully read Failure to do so could result in permanent damage to the unit Particular attention should be give...

Page 2: ...uct can be made unless they have been expressly approved by Quantum Cryogenics as this could void the user s authority to operate the equipment ATEX Hazardous Area Compatibility Ensure that the gauge is suitable for use in the environment of the hazardous zone This should be done by reference to the ATEX code of the DigiTemp which is also marked on the face of the instrument and stated on the cert...

Page 3: ...e substances please contact Quantum Production Ltd for further advice INSTALLATION Installation should only be undertaken by qualified personnel familiar with safety practices and standards relating to intrinsically safe electrical equipment OPERATING INSTRUCTIONS The DCT150M W samples the temperature measured at the sensor once every minute In practice due to the slow change in the temperature of...

Page 4: ... and user must take measures to avoid accumulation of static charge on the non conductive enclosure surface Please refer to instructions in cleaning section below No changes or modifications to the product can be made unless they have been expressly approved by Quantum Cryogenics as this could void the user s authority to operate the equipment MOUNTING THE SENSOR In order to monitor the authentic ...

Page 5: ... welded to the tank without using undue force and then be held firmly in position by the use of a setting high thermal conductivity silicone sealant Welds Tank Surface Saddle Clips Sensor Housing Flexible Lead Fig 1 ii Alternatively the sensor housing may be fixed to the outside surface of the tank by inserting it into a stainless steel tube of suitable diameter which has been pre welded to the ta...

Page 6: ...e dimensions of the DCT150M Digital Tank Thermometer are shown in Fig 3 below The case is designed to be flush fitted in to a circular hole of diameter 4 inches 122 mm in the outer skin of the tank and to be fixed in position by means of rivets or screws Six fixing holes Ø5 mm are provided around the fixing flange of the unit on a PCD of 135mm The thermometer flange should be sealed to the outer s...

Page 7: ...e material and when installing great care must be taken not to damage it When fitting the sensor pay particular attention to the following points 1 Do not pull tug or stretch the sensor cable and ensure that the lead is not resting or snagged against a sharp edge or trapped between two surfaces 2 If the lead is too long it can be coiled and held together by a high temperature tie wrap making sure ...

Page 8: ...y cloth Strong solvents should not be used to clean the flange and front face MAINTENANCE The DCT150M W does not require any user maintenance other than cleaning as indicated above The face of the unit is permanently sealed to the enclosure and should not be removed Refurbishment and replacement batteries can be undertaken by returning the unit to Quantum Production Ltd WARRANTY The DCT150M W is g...

Page 9: ...ay 4 Digits 7 segment LCD optional Fahrenheit F configuration Resolution 1 C Dimensions ø150 40 mm rear cut out ø100 mm Fittings 6 fixing holes ø5 mm equi spaced on a 135 mm PCD Enclosure Glass fibre reinforced nylon with fully integrated sealed polycarbonate front face Sensor enclosed in ø6 mm x 50 mm sheath stainless steel Ingress Protection IP67 Serial Number Displayed on front face with year m...

Page 10: ...IEC 60079 0 2017 EN60079 11 2011 Ex ib IIC T4 T3 Gb Ex ib IIIC T135 C T184 C Db HAZLOC E115327 IEC EN UL CSA C22 2 no 60079 0 IEC EN UL CSA C22 2 no 60079 11 CSA C22 2 no 213 UL 121201 UL CSA C22 2 no 61010 1 Class I II III Div 2 Gps A D FG Nonincendive Non incendiaire Class I Zone 1 AEx ib IIC T4 T3 Gb Zone 21 AEx ib IIIC T135 C T184 C Db Ex ib IIC T4 T3 Gb Ex ib IIIC T135 C T184 C Db EMC ETSI 30...

Reviews: