Page 24 of 28
Operation
2.11.5 Output Data Format
Output data is in floating point format as illustrated in Table 2-4 below. A data string is
preceded by R if display is resistance value and P if display is % deviation.
NOTE
A negative value is possible on the lowest range or ranges when test lead zeroing is
enabled.
Table 2-4: Output Data Format
Range
Measurement Results
(Resistance)
R Value (Resistance)
Over Indication
R1
R = - - . - - - m
R = 999999 m
R2
R = - - - . - - m
R = 999999 m
R3
R = - . - - - -
R = 999999
R4
R = - - . - - -
R = 999999
R5
R = - - - . - -
R = 999999
R6
R = - . - - - - k
R = 999999 k
R7
R = - - . - - - k
R = 999999 k
R8
R = - - - . - - k
R = 999999 k
R9
R = - . - - - - M
R = 999999 M
Range
Measurement Results
(%)
% Deviation
Over Indication
R1
P = - - - . - - %
P = 999999 %
R2
P = - - - . - - %
P = 999999 %
R3
P = - - - . - - %
P = 999999 %
R4
P = - - - . - - %
P = 999999 %
R5
P = - - - . - - %
P = 999999 %
R6
P = - - - . - - %
P = 999999 %
R7
P = - - - . - - %
P = 999999 %
R8
P = - - - . - - %
P = 999999 %
R9
P = - - - . - - %
P = 999999 %