Faults
MXD70 Suspended Solids / Turbidity
Setup and Operating Guide
- 37 -
Faults
E066
CH1
AUX mA Input Below 4mA
E116 CH2
The sensor input is less than 4mA, check sensor condition and connections. If the
message persists please consult with your supplier.
E166 CH3
E067
CH1
Sensor 0mV Cal Out of Spec
E117 CH2
The pH 0mV calibration for this channel is outside recommended specifications.
E167 CH3
E068 CH1 Calibration
Due
E118 CH2
The time since the last calibration was performed on this channel has exceeded the
time set in the calibration menu.
E168 CH3
E069 CH1 Planned
Service
Due
E119 CH2
The Planned Service interval for this unit has expired. Please contact Quadbeam
Techonlogies at the details below:
Quadbeam Technologies Ltd
10/16 Alpito Place
Pukekohe
Auckland
2120
New Zealand
Tel. +64 (09) 238 4609
Email [email protected]
NB. Quadbeam Technologies overseas users should contact their local distributor.
E169 CH3
E070
CH1
SD Card Checksum Error
E120 CH2
The SD Card store from which this channel was restored from has become
corrupted. Check the channel's settings and then save the settings again to the SD
card store.
E170 CH3
E071 CH1 Gain
Error
E121 CH2
The sensor gain has been exceeded. If this message persists, please consult with
your supplier.
E171 CH3
E072
CH1
Invalid Linearisation Curve
E122 CH2
A minimum of 2 linearisation points are required. Please check linearisation curve
settings in the channel setup menu for this channel.
E172 CH3
E073 CH1 Linearisation Over-Range
E123 CH2
The linearisation result is greater than 9999 (or 100.0% when using %).
E173 CH3
E074 CH1 Linearisation
Under-Range
E124 CH2
The linearisation result is less than 0.
E174 CH3
E075
CH1
Curve Low Limit
E125 CH2
The incoming probe signal is less than the lowest point in the linearisation curve.
E175 CH3
E076
CH1
Curve High Limit
E126 CH2
The incoming probe signal is greater than the highest point in the linearization
curve.
E176 CH3
E077 CH1 Custom
Error
E127 CH2
Contact your supplier for details.
E177 CH3
Summary of Contents for MXD70 Series
Page 1: ...MXD70 SERIES Multi parameter Monitor Suspended Solids Turbidity Setup and Operating Guide ...
Page 2: ......
Page 6: ...Contents 4 MXD70 Suspended Solids Turbidity Setup and Operating Guide Blank ...
Page 28: ...Appendix 26 MXD70 Suspended Solids Turbidity Setup and Operating Guide Appendix Blank ...
Page 49: ...Notes MXD70 Suspended Solids Turbidity Setup and Operating Guide 47 ...
Page 51: ......
Page 52: ......