Rev. 6-12
20
Programming Screen
How To Create a Program
The ult rasonic processor has t he abilit y to save up to 10 programs.
1. Select a program number f rom t he keypad, the program number w ill appear above t he
Amplit ude box.
For each of the following steps
: Af t er t ouching a f ield, the background w ill change to yellow
indicat ing an act ive f ield. Af t er ent ering data, t he ent ry is completed by t ouching Ent er (ENT),
t ouching the f ield a second t ime or by touching a new f ield.
2. Program a value int o the Amplitude f ield.
3. Program tot al Process Time (tot al act ive sonication t ime).
4. Program Pulse On t ime
– If no data is entered (00hr:00min:00sec), the unit w ill run
cont inuously w it hout pulsing.
5. Program Pulse Of f t ime
– If no data is entered, the unit w ill run continuously w ithout pulsing.
Process Time, Pulse On and Pulse Of f t imes are measured in Hours: Minut es: Seconds f ormat
.
6. Act ivate Temperature Monitoring (if necessary).
7. Program a Shut dow n Temperature (if necessary). Enter a temperat ure value.
During sonicat ion, if the temperat ure reaches t he Shutdow n value, t he unit w ill pause
sonicat ion. The unit w ill remain paused unt il the t emperature drops below t he pre-set value.
Once t he temperat ure drops below t he value, sonicat ion can be manually resumed.
8. Touch Save t o store the program t o t he memory number select ed in st ep 1.
9. Touch Run t o proceed t o t he follow ing screen.
10. Touch St art t o begin the program.
Program
Progress
Total time of active
sonication that has occurred
during the current program.
Rest or Off time is not
included in this value.
Total programmed time of
active sonication. Rest or
Off time is not included in
this value.