Page 11 of 25
QSA Global, Inc.
40 North Avenue Burlington, MA 01803
888.272.2242
781.272.2000
F: 781.359.9179
qsa-global.com
X-Ray Live Image Screen
Figure 7
X-Ray Live Image Screen
After 5 seconds, the system will emit X-rays, and transition to the
X-Ray Live Image
Screen
(Figure 7;
Settings
menu is
hidden by default). System Status Indicators remain amber-colored, and the buzzer pulse rate changes to 2-second
intervals.
To change settings while emitting X-Rays (i.e. on-the-fly), hold down the hat switch for 3 seconds (Figure 3) to show and
activate the
Settings
menu (see Figure 8).
Figure 8
Settings
menu re-activated
The
X-Ray Live Image
Screen
remains active until the red trigger is released, or 90 seconds has elapsed. When either
occurs, the following will happen:
System stops emitting X-Rays
Buzzer turns off