Loading the shaker rack with sample tubes that have
attached lids
Lids of sample tubes must be securely placed into the
slots at the edge of the shaker rack
Loading the shaker rack with sample tubes that have
screw-cap lids
Shaker rack plugs must be placed into the slots at the
edge of the shaker rack
6.
Press
Next
to proceed to starting the run or loading the centrifuge, depending on your
selected protocol.
WARNING
Risk of personal injury and material damage
[W5]
To prevent plastic crash, load the tubes properly. After a plastic crash, sharp
plastic particles could be inside the centrifuge. Be careful when handling
items inside the centrifuge.
Important
: Do not use 1.5 ml microcentrifuge tubes on the shaker. These microcentrifuge tubes can
cause filter-tips to stick during sample transfer. Using this tube on the shaker can damage the
pipetting system and can cause a centrifuge crash.