QIAcube Connect MDx User Manual 12/2020
134
6.
Plus signs (+) will be shown at different positions on the screen. For each of these, press the
center of the + sign. After one position is touched, the next position will be shown. The
graphic below shows the positions and the order in which the + signs will appear.
Positions and the order in which the plus signs will appear
7.
The calibration is finished after all five positions have been touched. Calibration results are
saved automatically. The previous screen is shown again.
8.
Press
Start
. The instrument continues initialization using the new calibration settings.
To cancel the calibration process, turn OFF the QIAcube Connect MDx.
7.3.5
O-Ring exchange
O-Ring replacement must be performed if the tightness test (see section 6.7.2, Tightness test) failed
or if the following issues are observed:
Uneven volume transfers
Dripping on the worktable
In any case, it is recommended to consult QIAGEN technical service. The replacement procedure
requires the O-Ring change tool and an O-Ring. See section 10, Appendix B - QIAcube Connect
MDx Accessories, for ordering details.