QIAcube Connect MDx User Manual 12/2020
122
During each cycle, an average summed dose rate of 28 to 46 mW*s/cm2 can be achieved by UV
LED light.
1.
To start the UV decontamination, press the
Tools
icon on the menu bar. Press the
Maintenance
tab and select the
UV Run
subtab. The screen shows the
Last Executed
UV run date and the
Cycle duration
.
UV run screen
2.
In the
Cycle
field, change the number of cycles. The default cycle number is 1 (approximately
12 minutes).
3.
Ensure that all disposable labware has been removed from the worktable.
Important
: Ensure that the waste drawer is closed. Do not open it during the UV run.
4.
Close the hood and press
Start
to begin the UV run.
5.
Press
Done
once the UV run is completed. The date of the last performed UV-run is updated
automatically.
WARNING
Risk of personal injury
[W29]
Do not expose your skin to UV-C light from the UV LED lamp.