qfix RapidHeat RT-2075-110 Quick Start Manual Download Page 2

2007842_EN_B

Qfix | 440 Church Road | Avondale, Pennsylvania 19311 USA |   +1 610-268-0585 |   +1 610-268-0588 |   [email protected] | www.Qfix.com

! WARNING ! FOLLOW ALL WARNINGS, PRECAUTIONS, AND INSTRUCTIONS FOR USE CONTAINED WITHIN THE PRODUCT GUIDE 

AND USER MANUAL FOR THE RAPIDHEAT™ OVEN.
! WARNING ! FOLLOW ALL WARNINGS, PRECAUTIONS, AND INSTRUCTIONS FOR USE CONTAINED WITHIN THE PRODUCT GUIDE 

AND USER MANUAL FOR AQUAPLAST RT® & FIBREPLAST® THERMOPLASTICS.
! WARNING ! THE FRAME AND THE THERMOPLASTIC MATERIAL MAY BE HOT! HANDLE WITH CARE. ALLOW TO COOL SLIGHTLY 

PRIOR TO PATIENT CONTACT TO AVOID PATIENT INJURY.
! WARNING ! THE BOLUS MAY BE HOT! HANDLE WITH CARE. ALLOW TO COOL SLIGHTLY PRIOR TO PATIENT CONTACT TO AVOID 

PATIENT INJURY.
! WARNING ! THE OVEN RACKS MAY BE HOT! HANDLE WITH CARE. USE HAND PROTECTION WHEN HANDLING.

AQUAPLAST RT® & FIBREPLAST® THERMOPLASTICS

Prior to taking the patient into the simulation room, turn on the RapidHeat™ Oven by pressing the power button. 

Manual Mode

1.  When the Program Indicator (P01, P02, etc.) stops flashing, press the INCREASE 

 or DECREASE 

 button to set the desired 

temperature. The light will flash next to the small TEMPERATURE   icon. Refer to table below.

2. Press the TIMER 

 button. The light will flash next to the small TIMER   icon. Press the INCREASE 

 or DECREASE 

 button to 

set desired heating time.

3. Press START/STOP 

START

STOP

 to begin the preheat cycle. The display will flash “PrH” for PREHEATING. Jump to “Heating the Mask or Bolus” 

section.

! NOTE ! When changing the temperature, verify the light next to the small TEMPERATURE

   

icon is flashing. When changing the 

heating time, verify the light next to the small TIMER

   

icon is flashing.

Program Mode

1.  While the Program Indicator (P01, P02, etc.) 

is flashing

, press the INCREASE 

 or DECREASE 

 button until the desired Program 

is selected. 

2.  Press START/STOP 

START

STOP

 to begin the preheat cycle.

! NOTE ! The display will indicate the last memorized program by flashing P01 (Program 1), P02 (Program 2), etc. If the display stops 
flashing before the desired program has been selected, press the P 

P

 button and press the INCREASE 

 or DECREASE 

 

button until the desired program has been selected.

Heating the Mask or Bolus

1.  The oven will beep and the display will indicate PrH READY when the oven has completed the pre-heating cycle and is ready to heat 

the thermoplastic mask or bolus.

2. Place the Aquaplast RT or Fibreplast mask or bolus on top of the mesh liner. 
3. Using the mesh liner as a transfer sheet, place the mesh liner onto the oven rack. 

! NOTE ! Place the bottom edge of the mask frame towards the back of the oven and should align to the back edge of the oven rack. 

The head of the mask should align to the front edge of the oven rack in the front of the oven. 

4. When the oven door is shut, press START/STOP 

START

STOP

 and the timer will begin counting down.

•  During the softening cycle, the actual temperature of the oven chamber can be observed by pressing the TEMPERATURE 

 

button once.

•  To increase or decrease the oven temperature, press the INCREASE 

 or DECREASE 

 button until the desired temperature is 

reached.

•  To increase or decrease the countdown timer, press the TIMER 

 button followed by the INCREASE 

 or DECREASE 

 

buttons to re-set desired heating time.

•  Press the LAMP 

 button to illuminate the oven interior. The light will remain on for 60 seconds.

•  Opening the oven door during the softening process will pause the timer. The timer will resume the countdown when the oven door 

is shut.

5. When the timer reaches zero, the oven will beep three times and then once every 60 seconds indicating the softening cycle has been 

completed. The display will indicate READY. Using the mesh liner as a transfer sheet, remove the mask or bolus from the oven rack. 

6. Ensure the mask is cool enough for patient comfort prior to patient contact. 

Recommended Heating Times and Temperatures for Qfix Thermoplastic Masks and Bolus

Thermoplastic

Heating Time

Heating Temperature

Aquaplast RT & Fibreplast Masks

3–11 minutes

74°C (165°F)

Aquaplast RT Bolus

10–15 minutes

74°C (165°F)

! WARNING ! DO NOT HEAT THE MASK OR BOLUS BEYOND 30 MINUTES.