QBC Diagnostics ParaLens Quick Start Manual Download Page 4

2

1

5

7

6

TYPE B

4

STANDARD INCLUDED EQUIPMENT:

3

p/n: 427751

p/n: 424273

p/n: 427711

p/n: 427708

p/n: 424278

p/n: 427707

p/n: 427710

ParaLens OptiLume 

Light Source
ParaLens Adapter w/ Objective
Power Pack w/ Adapters
Immersion Oil

1.

2.
3.
4.

ParaViewer
Screwdriver
Malaria Tube Starter Kit 

(Stored in compartment under 
Immersion Oil and ParaViewer)

5.
6.
7.

Also included (not shown)

OptiLume System Carrying Case 

w/ Foam Inserts (p/n 427712)
Literature Packet (p/n 427709)

1.

2.

PARALENS ADAPTER SYSTEMS:

INTRODUCTION:

The ParaLens is a complete epi-

fl

 uorescence microscopy system.  Contained 

within the system are the necessary objectives, 

fi

 lter, and beamsplitter 

for use during 

fl

 uorescent microscopy.  The standard con

fi

 guration of the 

objective is appropriate for 

 tube length microscopes.

,

CAUTION:

The ParaLens System uses a Class 2 LED product. Never stare into 

OptiLume Light Source or adapter when ParaLens is turned “ON”

FIGURE 1. ParaLens Adapter w/ Objective

Adapter Body
Threads
Objective
Light Input Shaft
Set Screw
Thumbscrew

1.
2.
3.
4.
5.
6.

6

2

5

1

3

4

1)  Thumbscrew
2)  Power pack 
     insertion point 
3)  Intensity knob

FIGURE 2. OptiLume Light Source

3

2

PARALENS INSTALLATION PROCEDURE:

The ParaLens Adapter – see 

fi

 gure 1 below – is equipped with 

a threaded end (Item 2) that turns 

freely within the ParaLens body 

(Item 1).  To install ParaLens 

onto a standard microscope 

begin by aligning the threaded 

end of the ParaLens Adapter 

with threaded screw holes in 

the microscope nosepiece.  Turn ParaLens Adapter clockwise to engage threads 

on microscope nosepiece.  Continue until ParaLens Adapter is securely fastened to 

nosepiece of microscope.
The ParaLens Adapter may be rotated to any desired position and secured at given 

location by tightening the set screw (Item 5) with provided ParaLens screwdriver.  It is 

not necessary to tighten the set screw for the ParaLens to function properly.
The ParaLens OptiLume Light Source must now be attached to the ParaLens Adapter.

OPTILUME LIGHT SOURCE INSTALLATION PROCEDURE: 

Begin by removing the ParaLens Adapter thumb screw (Item 6, Figure 1). 
Slide the ParaLens OptiLume Light Source onto the 

ParaLens Adapter. 
Tighten thumb screw (Item 1, Figure 2) on ParaLens 

Optilume Light Source to secure the OptiLume Light 

Source onto the ParaLens Adapter.
If using power pack, select proper wall adapter 

and connect to power pack by aligning the tabs on 

the adapter with the slot on the power pack. Push 

together and slide downward. Power pack and 

adapter are connected when a “snap” is heard.  
Plug power pack into ParaLens Optilume light source 

at insertion point (Item 2, Figure 2).
Plug power pack into an approved electrical 

outlet. 

Be sure that outlet is always within reach as line cord is the means of 

emergency disconnection.

The ParaLens Fluorescence Microscopy System should now be completely installed.  

A blue light should now be visible below the ParaLens objective.