Q-Tec 1000AD Instruction Manual Download Page 2

Instruction  Manual

AC/DC  Clamp  Meter

Model  No  :  1000AD

      We  thank  you  very  much  for  your  purchasing  AC/DC 

CLAMP  METER  1000AD. 

This  model  is  a  most  reliable,  high  precision  instrument, 

designed  by  our  excellent  technology.  Before  you  use 

your  new  instrument,  please  read  this  INSTRUCTION 

MANUAL  completely  and  familiarize  yourself  thoroughly 

with  all  functions  and  connections.

General  description



Display  :  3  3/4  digit  LCD,  max  reading  of  approx. 

4000.



Measuring  method  :  Dual  integration  method



Range  :  Auto  ranging.



Polarity  mark : Automatic. "-"  Polarity display.



Over-range  indication  :  [  OL  ]  mark  indication.



Auto  power  off  :  In  the  process  of  measuring,               

no  matter  the  function  key  or  switch,  if  there  is  no   

any  action  on  them  within  15  minutes,  the  meter 

will  be  "Auto  Power  Off”(Sleep  Mode)  to  save  the 

power.



Sampling  :  Reading  3times/second



Low  battery  indication  : 

  mark  is  display  when 

the  battery  voltage  drops  below  operating  voltage.



Power  supply  :  AAA  1.5V  ×  2pcs



Operation  temperature  :  0~40℃,  80%RH(max.)



Storage  temperature  :  -20℃~60℃,  70%RH(max.)



Size  &  Weight  :  190  X  56  X  32(H)mm

      Approx.  140g  (included  batteries  and  case)



Clamp  inside  diameter  :  Max.  45mm



Accessories  :  Instruction  manual, 

                              Test  probe,  Safety  soft  case

Mesurement  specification

DCA
ACA

Range

Resolution

Accuracy

400.0A

0.1A

±3.0%rdg

5dgt

1000A

1A

DC.V

Range

Resolu-

tion

Accuracy Input Impedance Max. input

400.0㎷ 100㎶

±1.3%
+2dgt

〉100㏁

600DC

or

600AC

rms

(Sine 

wave)

4.000V

1㎷

11㏁

40.00V

10㎷

10㏁

400.0V 100㎷

600V

1V

±2%+2dgt

AC.V

50/60

Hz

Range

Resolu-

tion

Accuracy Input Impedance

4.000V

1㎷

±2.3%

  +2dgt

11㏁

40.00V 10㎷
400.0V 100㎷

600V

1V

10㏁

Conti-

nuity

Check

Range

Resolu-

tion

Buzzer

sound

Test

Circuit

Input

Protection

400.0Ω 100mΩ

<50Ω

<0.7㎃

230Vrms

Clamp  Sensor

Warning  instruction  for  safe  use

To  ensure  that  the  meter  is  used  safely,  be  sure  to 

observe  the  instruction  when  using  the  instrument.

∎ 

Never  touch  on  the  case  that  exceed    1kVA.

∎ 

Pay  special  attention  when  measuring  the  voltage  of 

AC  30Vrms(42.2V  peak)  or  DC  60V  or  avoid  injury.

∎ 

Never  apply  an  input  signals  exceeding  the  maximum   

rating input voltage.

∎ 

Never  use  meter  for  measuring  the  line  connected 

with  equipment(i.e.  Motor)  that  generates  induced 

or  sure  voltage  since  it  may  exceed  the  maximum 

allowable  voltage.

∎ 

Never  use  meter  if  the  meter  or  test  leads  are 

damage  or  broken.

∎ 

Never  use  un-cased  meter.

∎ 

When  connecting  and  disconnecting  the  test  leads, 

first  connect  the  ground  lead  (black).  When 

disconnecting  them,  the  ground  lead  must  be 

disconnected  last.

∎ 

Always  keep  your  fingers  behind  the  finger  guards 

on  the  probe  when  making  measurements.

∎ 

Be  sure  to  disconnect  the  test  pins  from  the 

circuit  when  changing  the  function.

∎ 

Before  starting  measurement,  make  sure  that  the 

function  and  range  are  properly  set  in  accordance 

with  the  measurement.

∎ 

Never  use  meter  with  wet  hands  or  in  a  damp 

environment.

∎ 

Never  open  tester  case  except  when  replacing 

batteries.  Do  not  attempt  any  alteration  of  original 

specifications.

∎ 

To  ensure  safety  and  maintain  accuracy,  calibrate 

and  check  the  tester  at  once  a  year.

Description  of  panel

①  Function  Switch.

      This  rotary  switch  selects  the  measurement 

function  when  aligned  with  function  symbols  on 

the  panel.

②  Data  Hold  Button  :  This  button  is  used  to  hold 

data  during  a  measurement.  When  this  switch  is 

pressed,  the  measuring  data  is  held  and  D.H 

appears.  In  the  display,  the  display  remains 

unchanged  if  the  input  signal  changes.  Press  this 

button  once  again  releases  the  hold.

③  Liquid  Crystal  Display.

④  V,    Input  Terminal.

      Input  terminal  for  voltage  and    measurement.

⑤  COM  Input  Terminal.

      This  is  the  COMMON  input  terminal  for  voltage

      &    measurement.

⑥  Wrist  Strap

⑦  Zezo-Adj.  Button.

      This  button  is  used  the  DC  current  range  only.

      Once  this  is  pressed,  the  reading  value  shall  be 

set  to  zero  and  be  used  as  a  zero  reference 

value  for  other  subsequent  measurement.

⑧  Jaw  Opening  Lever

Measuring  Method

     

∎ 

AC/DC  Current  Measurement(Auto  ranging)

      Set  the  function  switch  to  "AC/DC  A"  range. 

      Clamp  only  one wire in the  circuit  to  be  measured.
      AC/DC  Current  measurement  is  limited  to  the  circuit 
      under  600V  AC/DC.

① When measuring DC current, Zero adj. should be  made.

②  Two  or  three  power  lines  at  once,  not  be  measured 

by  the  clamp.

③  The measured  value will be  displayed  on  the  display.

∎ 

MEASUREMENT  OF  DC  VOLTAGE(DCV)

      (AUTO-RANGING)

①  Set  the  function  switch  to  "DCV"

②  The  "DC"  mark  will  be  displayed  on  the  left  side 

in  the  display.

③  Connect  the  test  leads  to  the  circuit  to  be  measured, 

the  range  will  change  automatically  to  the  level  that 

will display  the input voltage with  the best resolution.

④  The  value  indicated  in  the  display  window  is  the 

measured  value  of  voltage  with  proper  decimal 

point  and  annunciator  indication.

∎ 

MEASUREMENT  OF  AC  VOLTAGE(ACV)

      (AUTO-RANGING)

①  Set  the  function  switch  to  "ACV"

②  The  "AC"  mark  will  be  displayed  on  the  left  side 

in  the  display.

③  Connect  the  test  leads  to  the  circuit  to  be 

measured,  the  range  will  change  automatically  to 

the  level  that  will  display  the  input  voltage  with 

the  best  resolution.

④  The  value  indicated  in  the  display  window  is  the 

measured  value  of  voltage  with  proper  decimal 

point  and  annunciator  indication.

∎ 

MEASUREMENT  OF  CONTINUITY  CHECK( )

①  Set  the  Function  switch  to  " ".

②  The  " "  mark  is  indicated  on  the  right  side  in  the   

display.  This  is  the  continuity  check  mode  in           

continuity  mode  a  buzzer  sounds  when  the             

resistance  of  the  circuit  to  be  measured  is  less   

than  approx.  50  Ohms.

Replacement  of  Battery

In  order  to  prevent  electrical  accidents,  when 
you  replace  the  battery,  the  function  switches 
off  and  the  test  leads  should  be  separated  from 
the  measuring  circuit.

①  When  the  batteries  becomes  exhausted  or  drops 

below  the  operating  voltage  the  "

"  mark  is 

displayed.

②  Turn  the  function  switch  "OFF"  and  remove  the 

test  leads  from  all  test  circuit,  prior  to  installing 

fresh  batteries.

③  Remove  the  battery  cover  at  back  side.  Insert  a 

tip  of  pincette  or  pen  in  the  hole  and  push 

downward.

④  Replace  the  two  batteries(AAA  size.  UM-4)  with 

new  ones  as  quickly  as  possible.  Then  inserting 

the  batteries.  observing  correct  polarity.

---------------------------------------------------------------

Q-TEC  ELECTRONIC  CO.

15-10,  Bokji-ro  120  beon-gil, 
Siheung-si,  Kyeonggi-do,  KOREA
429-807
TEL  :  (031)312-6213      FAX  :  (031)314-0049

http://www.qteckorea.com

Summary of Contents for 1000AD

Page 1: ...rms Sine wave 40 00V 10 400 0V 100 600V 1V 10 도통 체크 Range 분해능 도통 부저 음 Test Circuit Input Protection 400 0Ω 100mΩ 약50Ω이하 0 7 230V rms 각부의 명칭 및 기능 로터리 스위치 전원 On Off ACA DCA DCV ACV 선택 스위치 Data Hold 버튼 측정값을 화면상에 고정합니다 한 번 누르면 그때의 측정값이 고정되며 입력신호가 변하여도 표시된 측정값은 변하지 않습니다 Data Hold를 해제하기 위해서는 버튼을 다시 한 번 누릅 니다 액정 표시 부 3 3 4 Digit 액정 최대 유효표시 4000 입력단자 V 측정 시 적색 테스트리드 접 속 단자 Com 입력단자 V 측정 시 흑색 테스트리드 접속 단자 핸...

Page 2: ...alibrate and check the tester at once a year Description of panel Function Switch This rotary switch selects the measurement function when aligned with function symbols on the panel Data Hold Button This button is used to hold data during a measurement When this switch is pressed the measuring data is held and D H appears In the display the display remains unchanged if the input signal changes Pre...

Page 3: ... 流電圧 直流電圧 導通チェックがそれぞれ設定でき ます また電源をオン オフするスイッチでもありま す OFF の位置で電源が切れます Data Hold ボタン 測定中のデータを保持するスイッチです 1回押すとその ときのデータは保持され 以後 入力信号が変化しでも表 示は変わりません データホールドの解除は 再度ス イッチを押して下さい 液晶表示部 入力端子 電流測定以外の全てのプラス側入力端子で 付属の赤 のテストリドを接続します COM端子 電流測定以外の全てのマイナス側 共通COMMON 入力 端子で 付属の黒のテストリードを接 続します ハンドストラップ 手首などに通して使います ZERO ADJ Button 直流電流レンジにおいて 0 調整を行うためのButtonで す 入力が無い状態でZERO ADJ Buttonを押して 0 点 調整を行います 開閉レバー 電流測定の...