Q Imaging wLS Quick Installation Download Page 2

TROUBLESHOOTING

The wLS is a relatively simple system and is easy to operate. The following is a checklist should the product 
does not operate as expected. 

Unit does not power up — nothing appears on the pod screen. 

Is the power jack fully inserted in the wLS head? Is the mains lead fully inserted in the DC power 
supply? Is the mains supply switched on?

Display does not respond to any button presses.

 

Is the pod connector fully inserted in the wLS head? Was the pod connector inserted after power up? 
If so, power down and power up again once pod connector is fully inserted to reboot the system. 

LEDs fail to switch on with message on screen “Hot Lamps.” 

LEDs are overheating so check that there is sufficient clearance for airflow around the wLS head. 
Check that the ambient temperature is below the maximum operating temperature. Check that there 
are no local heat sources close to the head. Listen for the fan switching on. It should not run 
continuously, only when deemed necessary by the system's thermal management controls. If the fan 
does not operate with all the LEDs on and set at maximum intensity, then there is an internal fault 
and the unit must be returned to QImaging for repair. 

Illumination appears to be weak.

Are the intensity settings on the pod turned up? Has the single optical setup procedure for the wLS 
been followed? Is the microscope set up correctly? Check for shutters and apertures being open and 
for appropriate filters and cubes. If microscope previously used a mercury or metal halide lamp, check 
for damage or frosting of optics in light path. 
 

The Illumination is not uniform over the field of view. 

If the center of the field-of-view is brighter than edges, readjust the focus lens (Step 3 previous page)
to ensure homogenous illumination. If the illumination appears off center, carefully slacken off the 
fixing retaining the wLS onto the epi-port and rotate the head while viewing the field of view. If the 
illumination offset follows the rotation, then the optics within the wLS have lost their alignment, 
possibly through mishandling and the product will need to be returned to QImaging for realigning. If 
the illumination offset remains unchanged when rotating the light source, then the problem is within 
the microscope and a person competent with microscope servicing will need to investigate the cause 
of the problem.

www.qimaging.com 

©2015 QImaging. All rights reserved.

58-517-002 DOC;wLS QUICK INSTALLATION GUIDE Rev A0 

03-0027 Rev A3

Contact QImaging Customer Service for additional information: 1-800-874-9789 or www.qimaging.com

ROUTINE CARE

AND MAINTENANCE

 

The wLS requires little or no 
maintenance throughout its 
life. There are no field 
serviceable parts to there is 
no need to remove the 
covers. Cleaning of the 
external surfaces can be 
carried out with a mild soap 
and water used to lightly 
dampen a lint-free cloth. 
Ensure that no liquid is 
allowed to enter the 
product through vents and 
panel edges. Avoid optical 
surfaces. Cleaning of optical 
surfaces maybe necessary if 
debris or fingerprints 
accidently come into 
contact with the lens during 
installation. In the first 
instance remove any loose 
debris with an air duster 
(aerosol or rubber blower). 
Fingerprints or other liquid 
type contaminants should 
be removed using standard 
lens cleaning procedures. 
Do not flood the lens 
surfaces with fluid as liquid 
could enter the product and 
cause damage.

Reviews: