MEASUREMENT
Cycle
Slope
Exit
FLUORESCEIN CALIBRATION
Insert
DI
Solution
Click
Zero
Button
Insert
50
ppb Solution
√
Click
Slope
Button
Press cycle to choose solution
MEASUREMENT
Cycle
Continue
Exit
FLUORESCEIN CALIBRATION
Insert
DI
Solution
Click
Zero
Button
Insert
50
ppb Solution
√
Click
Slope
Button
√
Calibration Succeed
Press Continue to Proceed