NOTE
*
The nominal capacity provided by the tank manufacturer may be greater than the
maximum safe (net or effective) capacity that can be practically filled. Please keep this in
mind as you configure your LSP-X01 Series sensor for practical purposes.
9
Sensor Maintenance and Precaution
9.1
Methods to Cleaning LSP-X01 Series Transducer
For best performance, following the steps below to clean the submersible pressure transducer if necessary:
1. Hold the body of the transducer with one hand and carefully remove the protective nose cap by
simply unscrewing it from the transducer body. Do not touch the sensor diaphragm and do not dry
the inside portion of the transducer, as you risk damaging the pressure sensor.
2. Place the transducer in a vertical position with the sensing end facing downward in a bowl containing
Pyxis Probe Cleaning Solution for approximately 1 minute.
3. Place the protective nose cap in the same solution for approximately 1 minute.
4. Rinse the bowl with clean water and wipe dry the external casing of the transducer and the protective
nose cap.
5. Screw the protective nose cap back into place on the transducer body.
The Pyxis Probe Cleaning Solution is from the
Inline Probe Cleaning Solution Kit
(P/N: SER-01) which can
be purchased at our online E-Store
https://pyxis-lab.com/product/st-series-probe-cleaning-kit/
Figure 17.
Inline Probe Cleaning Solution Kit
LSP-X01 Series User Manual
| +1 (866) 203-8397
14