Specifications
Item
DW-2100P-US
DW-2100P-EU DW-2100P-US-G DW-2100P-EU-G
P/N
42135
42136
42137
42138
Turbidity Wavelength
Warm White
860 nm
Warm White
860 nm
Light Source
LED
Turbidity Dual Range
0.001 – 10 / 10 - 40.00 NTU
Turbidity Accuracy
+ 0.005 NTU or 2% <10NTU
Turbidity Repeatability
+ 0.001 NTU or 0.5% <10NTU
Free Chlorine Method
Bare Gold - Electrochemical Method
Free Chlorine Range
0.01 –5.00 ppm as FCl2
pH Method
Electrochemical Method
pH Range
0.01 – 14.00 pH
Measurement Accuracy
0.001NTU or ± 1% Full Scale
±0.01 ppm or 1% Full Scale w/pH compensation to 9.0
±0.01 pH
Minimum Resolution
0.001 NTU / 0.01 ppm FCl / 0.01 pH
Response time
4s – Turbidity /
T95≤60s
– FCl /
T95≤5s
- pH
Compliance
EPA-180.1/334.0 ISO-7027/7393 EPA-180.1/334.0 ISO-7027/7393
Measurement Interval
Continuous Measurement
Display
7-inch LCD Color Industrial Capacitive Touch Screen
Storage Capacity
Built-In 4GB of Ram for Storing up to 1-Million Data/Event Records
Power Requirement
96-260VAC / 50-60 Hz; 10A Fuse; 200 W
Output
3 x 4-20 mA / RS-485 Modbus - RTU / Modbus TCP
Input
3 x 4-20 mA / RS-485 Modbus - RTU
USB
1 x USB host, for data downloading and screen upgrade
Internet
RJ-45 socket, Modbus-TCP
Panel Operational Temperature
40 – 113
o
F (4-45
o
C)
Storage Temperature
Instrument: -4 – 131
o
F (-20 – 55
o
C) / Sensors 32 – 122
o
F (0 – 50
o
C)
Sample Water Temperature
40 – 104
o
F (4-40
o
C)
Sample Water Pressure
7.25 – 30 psi (0.05 – 0.2MPa)
FR-200 Sample Water Flow Rate
600-1,000mL / Minute
FR-200 Line Size
¼-inch Inlet /
⅜
-inch Overflow / 20mm Outlet / ½-inch Drain
Rating
IP-65 Panel-Display / IP-67 Sensors
Regulation
CE / RoHS
Relative Humidity
20% - 90% (No Condensation)
Altitude
<6,561 feet (<2,000 Meter)
Dimensions (HxWxD)
Panel (DW-2100P) 790H x 450W x 235D mm
Approximate Product Weight
DW-2100P ~ 20 kg
Pyxis 4G CloudLink
TM
NA
Included
CloudLink
TM
Bands
NA
Global B1/2/3/4/5/7/12/13/14/20/28/66/7
CloudLink
TM
Protocols
NA
IP/TCP/UDP/HTTP/HTTPS/Modbus
*NOTE*
- Pyxis Lab is consistently updating technologies, as such, specifications may change without notice.Technical specifications on the
LT-739/B and ST-765SS-FCL sensors can also be found in their respective Operation Manuals.Contact [email protected] for details or
Summary of Contents for DW-2100P
Page 1: ...Pyxis Lab Inc 1729 Majestic Drive Suite OS Lafayette Colorado 80026 www pyxis lab com...
Page 12: ...DW 2100P User service pyxis lab com 1 866 203 8397 12 Figure 2 FR 200 Connections...
Page 31: ...DW 2100P User service pyxis lab com 1 866 203 8397 31...
Page 37: ...DW 2100P User service pyxis lab com 1 866 203 8397 37 Figure 31 History Curve Screen...