8.2 Factory Reset
Use the following steps to restore all device parameters to factory default:
1. Power on the SP-200 by pressing
.
2. Press
Cal/Sys
(
) to launch the
Information
screen.
3. Press
Diag
(
) to launch the
Diagnosis
screen.
4. Allow 5–10 seconds for the message in the top-right corner of the display change from
Starting BTLE...
to
BTLE Started
.
5. Press
Factory
(
). The display updates as shown in Figure 20.
6. Press
OK
(
) to start the factory reset.
7. The message “Data recovery is successful!” will appear on the display (Figure 21).
Figure 20.
Figure 21.
SP-200 User Manual
| +1 (866) 203-8397
14