6.5
Diagnosis Screen
After the device has been calibrated and installation has been completed, to check diagnosis, click on
Diag-
nosis
. When in the
Diagnosis
screen you can view the Diagnosis Condition of the device. This feature may
be used for technical support when communicating with
To preform a Cleanliness Check, first select the
Diagnosis Condition
which defines the fluid type that the
ST-500 Series sensor in currently measuring, then click
Cleanliness Check
. If the sensor is clean, a green
Clean
message will be shown. If the sensor is severely fouled, a red
Dirty
message will be shown. In this
case, follow the procedure in the
Methods to Cleaning the ST-500 Series Sensor
section of this manual.
Figure 16.
7
Outputs
7.1
4–20mA Output Setup
The 4–20mA output of the ST-500 and ST-500SS sensor is scaled as:
• PTSA:
–
4 mA = 0 ppb
–
20 mA = 300 ppb
The 4–20mA output of the ST-500RO sensor is scaled as:
• PTSA:
–
4 mA = 0 ppb
–
20 mA = 40 ppb
7.2
Communication using Modbus RTU
The ST-500 Series sensor is configured as a Modbus slave device. In addition to the ppb PTSA value, many
operational parameters, including warning and error messages, are available via a Modbus RTU connection.
Contact Pyxis Lab Customer Service (
) for more information.