background image

8. Once the Cleanliness Check is completed a

Clean

message (Figure 12) or

Sample cell fouled

mes-

sage (Figure 13) will appear towards the bottom of the display.

9. Cleanliness check is now complete. Press

Exit

(

) to return to measurement mode.

Figure 11.

Figure 12.

Figure 13.

8.2

Bluetooth Connection to Devices

The SP-380P uses a built-in Bluetooth Low Energy Connection (BTLE) to connect wirelessly to a smart phone
via the

uPyxis®

Mobile App or to a computer via the included Bluetooth Adapter (P/N: MA-NEB) and the

uPyxis®

Desktop App. To allow the SP-380P to connect via Bluetooth with other devices follow the steps

below:

1. Power on the SP-380P by pressing

.

2. Press

Info

(

) to launch the

DEVICE INFORMATION

screen.

3. Press

Diagnosis

(

or

) to launch the

SYSTEM DIAGNOSIS

screen.

4. Allow 5–10 seconds for the message in the top-right corner of the display change from

Starting BTLE...

to

BTLE Started

(Figure 9).

5. Choose to connect via one of two options:

(a) The

uPyxis®

Mobile App (see the

Use with uPyxis® Mobile App

section), or

(b) The

uPyxis®

Desktop App (see the

Use with uPyxis® Desktop App

section).

8.3

Factory Reset

Use the following steps to restore all device parameters to factory default:

1. Power on the SP-380P by pressing

.

2. Press

Info

(

) to launch the

DEVICE INFORMATION

screen.

3. Press

Diagnosis

(

or

) to launch the

SYSTEM DIAGNOSIS

screen.

SP-380P User Manual

[email protected]

| +1 (866) 203-8397

9

Summary of Contents for 50402

Page 1: ......

Page 2: ...Handheld Fluorometer User Manual October 20 2020 Rev 2 00 Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com 2017 Pyxis Lab Inc Pyxis Lab Proprietary and Confidential...

Page 3: ...libration Two Point with Zero 7 8 Device Information and Diagnosis 8 8 1 Sample Cell Cleanliness Check 8 8 2 Bluetooth Connection to Devices 9 8 3 Factory Reset 9 9 Use with uPyxis Mobile App 11 9 1 D...

Page 4: ...elivery All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative or des ignate Pyxis warrants these services fo...

Page 5: ...iated with the cuvette are eliminated Large color graphic screen that can be read in direct sunlight 2 Specifications Table 1 SP 380P Specifications Item Specification Part Number P N 50402 Fluorescen...

Page 6: ...s 100 ppb PTSA Calibration Standard 500 mL 21001 Pyxis 200 ppb PTSA Calibration Standard 500 mL 21031 Pyxis 10 ppm Fluorescent Polymer Calibration Standard 500 mL 21054 Pyxis 20 ppm Fluorescent Polyme...

Page 7: ...and negative terminal signs in the compartment bottom snap four new AA alkaline batteries firmly into the battery holder 4 Replace the battery compartment cover and ensure that the sealing O ring is l...

Page 8: ...e the sample water to rinse the sample cell at least three times Allow 5 10 seconds for the SP 380P to stabilize Figure 2 For a sample containing 10 ppm Fluorescent Polymer the measured Fluorescent Po...

Page 9: ...the zero blank calibration 6 If the calibration succeeds a checkmark and instructions for the slope calibration will appear Figure 4 7 Rinse the sample cell three times with the desired Fluorescent Po...

Page 10: ...zero point 6 If the calibration succeeds a checkmark and instructions for the slope calibration will appear Figure 7 7 Rinse the sample cell three times with the desired PTSA standard Fill the sample...

Page 11: ...n HST Heavy fouling will pre vent the light from reaching the sensor resulting in inaccurate readings It is suggested that the SP 380P be checked for fouling and cleaned on a monthly basis Heavily con...

Page 12: ...evices follow the steps below 1 Power on the SP 380P by pressing 2 Press Info to launch the DEVICE INFORMATION screen 3 Press Diagnosis or to launch the SYSTEM DIAGNOSIS screen 4 Allow5 10secondsforth...

Page 13: ...igure 15 appear and the user can choose one of three options a Press to start the factory reset or b Press Cancel to return to the HELP screen or c Press Exit to abandon the factory reset entirely 7 A...

Page 14: ...e phone Do not pair the phone Bluetooth to the SP 380P Open uPyxis Mobile App Ensure that the SP 380P is discoverable by following the steps in the Bluetooth Connection to Devices section Once the app...

Page 15: ...well as update the firmware of the SP 380P by pressing Check Update If a firmware update is available press Get Firmware Once the new firmware is downloaded press Upgrade Firmware NOTE The firmware up...

Page 16: ...Net Framework 4 5 if not previously installed on the PC the USB driver for the USB Bluetooth adapter MA NEB the USB RS485 adapter MA 485 and the main uPyxis Desktop application Double click the uPyxis...

Page 17: ...Desktop App will default to the System screen From the System screen users can upgrade the firmware by selecting an appropriate firmware file contact service pyxis lab com for these firmware files and...

Page 18: ...og List and selecting the desired datalog these are separated by month The SP 380P will then populate any relevant log event from the selected datalog which can be viewed in more detail by clicking Re...

Page 19: ...hout any further key presses if it has not powered off If there are no key presses for 20 seconds the screen will darken and after another 20 seconds without key activity will power down The press of...

Page 20: ...onfirm that the SP 380P is clean Repeat the process as needed until the Cleanliness Check shows Clean 11 3 Storage Do not expose the SP 380P to an extreme high or low temperature condition such as lea...

Reviews: