background image

 

8.3 What is morning hypertension (morning surge)?   

Morning high blood pressure or morning surge is defined as the weekly average for morning 

blood pressure reading measured within 1 hour to 2 hours after awakening in the morning and 

exceeding 135/85mm Hg. Studies have shown that exaggerated morning blood pressure surge is a 

risk for cardiovascular events which includes ischemic and hemorrhagic stroke. Cardiovascular 

events have been shown to be exaggerated in the morning to coincide with morning high blood 

pressure. In fact heart attack, stroke and heart failure have been shown to fall particularly on a 

Monday amongst all the other days of the week. 

Organ damage and diabetic complications have also been shown to be linked with morning blood 

pressure surges just in the same way as small artery disease and multiple celebral infarcts in elder 

members of society. Morning high blood pressure has shown some correlation with initial stage 

and progression of atherosclerosis. Patients with well controlled blood pressure may still have 

high morning blood pressure and this happens in 50% of the cases. Patients with morning 

hypertension have a 78% more chance of stroke compared with 48% of other hypertensive 

patients without morning high blood pressure. Morning hypertension has also been associated 

with changes in heart size and rhythm. This may lead to heart attack or heart failure. 

Morning Hypertension can only detect within 1 hour to 2 hours after awakening, recommended 

user monitor their own blood pressure at home.   

 

 

Reference Standard

 

·

IEC  60601-1

2005

 

Medical  electrical  equipment-Part1:  General  requirements  for  safety  and 

essential performance. 

·

IEC  60601-1-2:2007  Medical  electrical  equipment  –  Part  1-2:  General  requirements  for  basic 

safety  and  essential  performance  –  Collateral  standard:  Electromagnetic  compatibility  – 

Requirements and tests 

·

EN  1060-1:1995+A2:2009  Non-invasive  sphygmomanometers  –  parts1:  General 

requirements 

·

EN  1060-3:1997+A2:2009  Non-invasive  sphygmomanometers  –  parts3:  Supplementary 

requirements for electro-mechanical blood pressure measuring systems. 

·

ANSI/AAMI  SP-10:2002+A1:2003+A2:  2006/(R)2008  Manual,  electronic,  or  automated 

sphygmomanometers 

·

ANSI/AAMI/ISO 81060-2-2009 Non-invasive sphygmomanometers-Part 2:Clinical validation of 

automated measurement type 

 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for PHBPB16TL

Page 1: ...ant Noted 6 4 2 Fitting the cuff 6 4 3 Body posture during measurement 7 4 4 Take Measurement 7 4 5 Data Transmission 8 4 6 Memory 9 5 Error Indication 9 6 Trouble Shooting 9 7 Specification 10 8 Abou...

Page 2: ...o vibration such as ambulances and emergency helicopters A location where there is gas or flame A location where there is water or steam A location where chemicals are stored A location where the unit...

Page 3: ...h several trials of testing to ensure the measurement accuracy The end user should conduct a manufacturer recommended inspection and calibration annually Blood pressure measurements determined with th...

Page 4: ...Cuff Type BF Applied Part Model PHBPB16TL Applicable Arm Circumference 8 6 to 14 inches Note Symbol for TYPE BF APPLIED PART Display...

Page 5: ...r demonstration purpose It is possible these batteries will therefore not last for 200 measurements The battery life can be confirmed in the bottom right of the display If the low battery symbol is di...

Page 6: ...at can raise blood pressure Try not to take your blood pressure if you are nervous or upset If you are nervous anxious or agitated your blood pressure will rise Rest for 5 10 minutes before taking a r...

Page 7: ...ensure the measurement accuracy 4 3 Body posture during measurement Relax place the elbow on the desk with palm facing up cuff should be at the heart level Fig 9 The reading accuracy may be affected...

Page 8: ...ic pressure is greater or equal to 90mmHg as abnormal The display will flash to remind the end user of the blood pressure abnormality If the device caused any discomfort during measurement or fail to...

Page 9: ...e memory mode and turn off 5 Error Indication List of Error code Error Cause How to correct Er 30 Power on cuff inflation rate is too low or main unit does not connect with the cuff 1 Reconnect the ai...

Page 10: ...during the day which also will affect by the human emotional and physical condition 2 Record the variance and consult to the doctor If the above suggestion doesn t remediable please dial Pyle Technic...

Page 11: ...Important information regarding Electro Magnetic Compatibility EMC With the increased number of electronic devices such as PC s and mobile cellular telephones medical devices in use may be susceptible...

Page 12: ...oltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable...

Page 13: ...ions equipment and the EQUIPMENT or SYSTEM The EQUIPMENT or SYSTEM is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of th...

Page 14: ...axed person and still have HBP Let s look at the facts about blood pressure so you can better understand how your body works and why it is smart to start protecting yourself now no matter what your bl...

Page 15: ...morning blood pressure and this happens in 50 of the cases Patients with morning hypertension have a 78 more chance of stroke compared with 48 of other hypertensive patients without morning high bloo...

Page 16: ...ymbol for manufacturer 0123 Symbol for CE Symbol for Follow operating instructions Symbol for TYPE BF APPLIED PART Symbol for electrical and electronic equipment IP21 Symbol for the IP classification...

Reviews: