JDX PVA
Revision H (2016)
16 of 39
Clean the Wetted Parts
3.8
1.
Clean the diaphragm with cotton tipped swabs and solvent.
Figure 13: Clean the Diaphragm
2.
Remove the feed tube from the nozzle plate and discard it.
3.
If necessary, soak the nozzle plate in ultrasonic cleaner for 5 minutes.
4.
Clean the nozzle plate with cotton tipped swabs and solvent.
Figure 14: Clean the Nozzle Plate
5.
Examine the parts for wear, replace if necessary.