12. Appendix
12.1 Set Point Limits and Defaults
12.1.1
System Operations
Description
Minimum
Value
Maximum
Value
Default
Value
Unit of
Measurement
High Pressure Limit –
Turn Compressor OFF
0.20
8.00
1.50
PSI
Low Pressure Limit –
Turn Compressor ON
0.10
7.90
0.30
PSI
12.1.2
Alarm Set Points
Description
Minimum
Value
Maximum
Value
Default
Value
Unit of
Measurement
Shutdown
High Pressure Alarm
0.20
8.00
3.00
PSI
Low Pressure Alarm
0.10
7.90
0.15
PSI
High Cabinet
Temperature Alarm
90
140
120
Deg F
YES
Low Cabinet
Temperature Alarm
34
42
34
Deg F
YES
High Humidity Alarm
6
18
7
%
YES
High Duty Cycle Alarm
0
99
50
%
12.1.3
Network Setup
Description
Format
Example
IP Address
Numeric (xxx.xxx.xxx.xxx)
192.168.001.100
Subnet Mask
Numeric (xxx.xxx.xxx.xxx)
255.255.255.000
Gateway Address
Numeric (xxx.xxx.xxx.xxx)
192.168.001.001
SNMP Trap Server
Address
Numeric (xxx.xxx.xxx.xxx)
192.168.001.255
Current Date/Time
Numeric (mm/dd/yy hh:mm)
01/27/11 10:02
Device ID
Alphanumeric (10 digits)
ABCD 12345
Keyword
Numeric (6 digits)
123456 (default)